background image

Standard Frequency Series Generators

Maintenance

16

M-2008-7-15

7.8 TROUBLESHOOTING GASSY X-RAY TUBE

SYMPTOM

MOST COMMON PROBLEM

HOW AND WHERE TO TEST

HIGH MA

 (continued)

Gassy X-Ray Tube

Gassy X-Ray Tubes are the result of three actions. The normal 

evaporation of filament and anode material over the normal 

life  of  an  X-ray  Tube,  accelerated  evaporation  of  filament 

and breakdown of anode material caused by exceeding tube 

limits,  and  immediate  breakdown  in  new  tubes  caused  by 

defective material or workmanship. The latter is pre-dominate 

in causing extensive damage to High Voltage Rectifiers and 

High  Voltage  Winding  when  the  defect  is  not  detected  and 

corrected quickly. All these actions result in gasses becoming 

active inside the vacuum tube. This gas is impure and allows 

accelerated movement of electrons. The pre-dominate cause 

is usually the anode run speed is too slow. See “

Anode Run 

Speed Too Slow

.”

•  This  gas  is  impure  and  allows  accelerated  movement  of 

electrons.  As  KVP  is  increased  MA  begins  to  increase. 

Space Charge cannot be changed to correct and stabilize 

mA. How radically a tube reacts to the set level of MA and 

kVp  is  determined  by  how  gassy  the  tube  may  be.  For 

instance, some tubes react quite well up to 200 MA and 

100 kVp, while others react quite violently when the tube 

is exposed above 50 kVp at 50 MA. Tubes may conduct 

properly when cold, but become unstable when they warm 

up. There is no one set way to measure how gassy a tube 

has become. 

If a tube is only slightly gassy, it must be 

replaced  because  it  will  cause  extensive  damage  to 

Rectifiers and to the High Voltage Winding.

 To test for 

slightly gassy X-ray Tubes, decrease KVP until the current 

stabilizes. The only sure way to test an extremely gassy 

X-ray Tube is by replacing the X-ray Tube and both High 

Voltage Cables.

Defective High Voltage Cable.

•  Shorting High Voltage Cables may cause higher MA’s or 

kicking the circuit breaker. Defective High Voltage Cables 

are  the  result  of  breakdown  of  insulating  material  that 

eventually leads to a short between wires or ground. This 

short allows a path for current to flow to GROUND. During 

a partial short the cable will cause symptoms much like a 

gassy X-Ray Tube. The result is higher MA.

In most cases the cable will short completely pegging the MA 

Meter. In this case there will not be output from the X-Ray 

Tube.  Shorting  X-Ray  Cables  if  left  uncorrected  can  cause 

opening of High Voltage Rectifier and High Voltage Winding 

in the Transformer.
Most  breakdowns  occur  at  the  receptacle  ends,  where  the 

cable enters the plastic and is bound by a pressed ring and 

at 90 bends. Also pinching, twisting, puncturing or allowing 

cables to become saturated with oil or other chemicals will 

break  down  insulation. 

To  test,  replace  the  High  Voltage 

Cables.

Содержание 325D

Страница 1: ...MEDICAL SYSTEMS Technical Publication M 2008 7 15 Maintenance Manual For Maintaining and Troubleshooting STANDARD FREQUENCY SERIES GENERATORS Contents Models TXR 325D TXR 325M TXR 425 TXR 525SFQ...

Страница 2: ......

Страница 3: ...sibility of any persons carelessly unwisely or unknowingly exposing themselves or others to X radiation It is important that everyone working with X radiation be properly trained and take adequate ste...

Страница 4: ...ycle of the equipment or system comes to an end becomes dangerous and need to be considered as harmful waste according to the international domestic and local regulations The manufacturer recommends t...

Страница 5: ...ng advisory symbols will be used throughout this manual Their application and meaning are described below DANGER ADVISES OF CONDITIONS OR SITUATIONS THAT IF NOT HEEDED OR AVOIDED WILL CAUSE SERIOUS PE...

Страница 6: ...g documents Ionizing radiation Type B equipment Dangerous voltage Ground This symbol indicates that the waste of electrical and electronic equipment must not be disposed as unsorted municipal waste an...

Страница 7: ...4 4 6 MA Stabilizer 4 5 TROUBLESHOOTING TECHNIQUES 5 5 1 Establishing Troubleshooting Techniques 5 5 1 1 Essentials Required for Producing X ray 5 5 1 2 Three Key Indicators for Effective Troubleshoot...

Страница 8: ...7 8 Troubleshooting Gassy X ray Tube 16 17 7 9 Troubleshooting Low MA 18 7 10 Troubleshooting No MA 18 20 8 COLLIMATOR 21 8 0 Theory of Operation 21 8 1 Mechanical Operation SID Selection 21 8 2 Elec...

Страница 9: ...2008 7 13 M 2008 7 14 and M 2008 7 15 Schematics are located in manual M 2008 7 13 1 1 TOOLS AND TEST EQUIPMENT REQUIRED FOR SERVICING X RAY SYSTEMS The following hand tools and products are required...

Страница 10: ...ements in bold lettering throughout this manual WARNING You are to have proper instrumentation to measure all energies used in and emitted from this equipment At no time are you to expose a person to...

Страница 11: ...ched to Bucky ON change the switch to Bucky OFF and retest If exposure meets condition of 5 above trouble is in the bucky circuit 7 If there is still no exposure initiated under condition 6 the proble...

Страница 12: ...of the Transformer so it can be stepped up to high voltage and Rectified for use by the X ray Tube 4 3 HIGH VOLTAGE TRANSFORMER High Voltage Transformers do not regulate current or voltage Its sole pu...

Страница 13: ...se provides the Keys to establishing effective troubleshooting Techniques The indicators are 1 High MA which can cause light images dark images and kicking circuit breakers 2 Low MA which mainly cause...

Страница 14: ...gh voltage secondary Because of loading kVp will decrease at the X Ray tube The bucking voltage selected from the 95 110 VAC supply bucks H1 to H2 Voltage The final Voltage difference is coupled to th...

Страница 15: ...readings are from the X ray Timer s software 6 3 MA METER The MA meter is a simple Digital resistor diode Display that indicates MA stations selected This meter cannot be adjusted 6 4 TESTING HIGH VO...

Страница 16: ...across major to minor line strap Should read 200 278 VAC Control Energizes but LED s are not lit Open fuse or loose fuse holder A1 or A2 With Line to Auto Strapped properly and with the Circuit Breake...

Страница 17: ...sert Check continuity of tube stator You should find continuity between any pair of the three 3 terminals Gently shake Tube Head or look into port to see if X ray Tube is BROKEN If the tube is broken...

Страница 18: ...posure is now possible Replace the EXPOSURE SWITCH TIMER SECTION 4 During Timer Standby continuous 120 VAC is required on Timer Pins J1 2 and J1 4 Place volt meter between those pins and confirm volta...

Страница 19: ...ed to the Bucky Motor Circuit B3 to B4 The Bucky Motor begins to drive and move the x ray filter grid across the front of the receptor for as long as the switch is depressed At some point the x ray fi...

Страница 20: ...en VISUAL INSPECTION Remove Collimator Select large and then small X ray filaments See if each filament lights up during PREP If both filaments light up properly the problem is in another area Disrega...

Страница 21: ...se EXPECTED VOLTAGE MEASUREMENTS WHEN A SHORT CIRCUIT EXISTS 2 Measure voltage from XC to XS or XC to XL With Prep Switch depressed voltage should be 100 200 VAC If it is remove the collimator and obs...

Страница 22: ...t is or is not operating properly SIMULATED LOAD TEST 1 Remove P1 P2 power cord from the Primary of the High Voltage Transformer 2 Place a Volt Meter between H1 and H2 terminals on the X Ray Control M...

Страница 23: ...ion of High Tension AC Current The Full Wave Bridge rectifies this current to DC DC current measurements dc milliamperes and are taken from and terminals on the Motherboard After the rectifier AC conn...

Страница 24: ...the tube is exposed above 50 kVp at 50 MA Tubes may conduct properly when cold but become unstable when they warm up There is no one set way to measure how gassy a tube has become If a tube is only s...

Страница 25: ...ake an exposure If MA is present you must replace High Voltage Rectifier Board You must establish the reason the Rectifiers shorted Normally a High Voltage Cable shorted or the X ray Tube became gassy...

Страница 26: ...dropped To test the High Voltage Rectifier Board and wiring remove the transformer from the tank and inspect all wiring and connections using a OHM Meter Open Space Charge Primary Transformer Open sp...

Страница 27: ...and the other on the S or L pin depending on which focus is selected Depress PREP and measure voltage If 15 16 VAC is present filament transformer is OK If voltage is not present check input voltage X...

Страница 28: ...move fuse 4 and measure resistance across the 2500 ohm resistor If the resistor is OPEN replace it NOTE If the 2500 ohm resistor heats up or opens you may have a short from MT 1 to MT 2 to a winding o...

Страница 29: ...are positioned manually with these knobs Back illuminated SID scales indicate all conventional film sizes for both 100 cm and 180 cm SID s as well as 40 inches and 72 inches in SIDs All Eureka Linear...

Страница 30: ...F3 on the power supply board and then to collimator receptical connections C and D 8 2 2 LIGHT FIELD LAMP CIRCUIT Refer to Schematic 70 08372 and Assembly Drawing 70 08371 This drawing is provided in...

Страница 31: ...tor Exposure Switch Prep Switch Earth Ground Earth Ground L1 L2 Connection Bucky Switch Timer Board 50 100 MA Adjustments 150 500 MA Adjustments SCR X Ray Tube Rotor Connections 07 08 09 Backup Contac...

Страница 32: ...se for planning or training BMET s MANUAL COST PER COPY Pre Installation Manual M 2008 7 12 20 00 Installation Manual M 2008 7 13 35 00 Operation Manual M 2008 7 14 25 00 Maintenance Manual M 2008 7 1...

Отзывы: