M-2008-7-15
Standard Frequency Series Generators
Maintenance
1
SECTION 1
INTRODUCTION
This manual is written as a guide to help Biomedical Technicians and X-ray Servicepersons become familiar
with TXR X-ray Control theory and circuitry. Those not trained in x-ray technology including electrical and
radiation safety should not attempt repair of x-ray equipment. When covers are removed, persons will be
exposed to Lethal voltages that exists throughout the entire circuitry of the equipment and they may be
exposed to cumulative doses of x-radiation.
This manual is one of a series of manuals provided with the x-ray system and should be used as a combined
approach to operating, testing, calibrating and validation by documenting results. The series of manuals
include: M-2008-7-12, M-2008-7-13, M-2008-7-14 and M-2008-7-15. Schematics are located in manual
M-2008-7-13.
1.1 TOOLS AND TEST EQUIPMENT REQUIRED FOR SERVICING X-RAY SYSTEMS
The following hand tools and products are required for the installation.
• Standard service engineers tool kit.
• Drill and assorted bits.
• Silicone Insulating Grease (proofing compound).
• Alcohol cleaning agent.
The following test equipment is required for Calibration and Service.
• Digital Multimeter.
• Non-invasive kVp Meter.
• Digital mAs Meter.
• Non-invasive Digital mR Meter
• Calculator.
• Stands or Test Instruments called for in sections of this manual.
Only for AEC purposes:
• Sensitometer.
• Densitometer.
Copper Plates for the Collimator Filter Holder (recommended for AEC calibration):
• 2 units of 1.0 mm thickness,
• 1 unit of 0.5 mm thickness,
• 2 units of 0.2 mm thickness,
• 1 unit of 0.1 mm thickness.
• Instead of Copper Plates it can be used Acrylic Plastic Plates:
• 6 units of 5 cm. thickness
• 5 units of 1 cm. Thickness
Only for Tomo purposes:
• Tomophantom tool.