Standard Frequency Series Generators

Maintenance

8

M-2008-7-15

SECTION 7

TROUBLESHOOTING CHARTS

7.1 ERROR CODE CHART

Models TXR 325 D, TXR 325 M, TXR 425 and TXR 525 SFQ X-Ray Controls come equipped with Self 

Diagnostics. Should a problem arise, observe the mAs meter display for one of the following error codes.

ERROR CODE

DESCRIPTION

O.L.

Indicates – Exceeds Tube Limit

E02

Indicates – KVP over range

E03

Indicates – KVP under range

E04

Indicates – Opening in MA cables or switch

E05

Indicates – Opening in Timing cable or switch

SCR

Indicates – Leaky or shorted SCR

TROUBLESHOOTING TECHNIQUE: THREE MAIN INDICATORS MUST BE OBSERVED

HIGH - LOW – NO MA

7.2  TROUBLESHOOTING THE X-RAY CONTROL AND METER CIRCUITS

SYMPTOM

MOST COMMON PROBLEM

HOW AND WHERE TO TEST

X-Ray Control will not energize

No power for X-ray L1, L2 

and ground

Check Power feeding X-ray including main panel. Requires 

240VAC Nominal. Range 

200-278 VAC

.

Circuit Breaker Defective

With Circuit Breaker in the on position check voltage across 

major to minor line strap. Should read: 

200-278 VAC

Control Energizes but LED’s are 

not lit. 

Open fuse or loose fuse 

holder 

A1

 or 

A2

.

With Line to Auto Strapped properly and with the Circuit 

Breaker on, check voltage from 

A1

 to 

A2

. Should be 

120-

122 VAC

.

No MA Meter Reading 

Open fuse or lose fuse holder 

F6.

With Line to Auto Strapped properly and with the Circuit 

Breaker on, check voltage from 

A1

 to 

F6

8 VAC

Defective MA Meter

Replace  MA  Meter,  Ribbon  Cable  Connector  or  MA 

Selector.

kVp Meter Reading not Stable 

– Numbers on the kVp Continues 

to countdown toward zero. 

Open Fuse or loose Fuse 

Holder 

F4 or F5

.

With Line to Auto Strapped properly and with the Circuit 

Breaker on, check voltage from 

A1 to F5 = 24 VAC. For 

Fuse 4 check voltage MT-1 to MT-2 = 100 VAC.

Major or Minor Selector 

defective.

Be sure that the kilovolt selectors are not stuck between 

contacts, but are properly indexed on a contact point. Make 

sure ALL contact points on both switches are good. 

KVP Meter Compensation 

Circuit

There may be an OPEN in the kVp Compensation Resistor, 

Compensation Transformer, or 

K1 – K7

 Relays are NOT 

ENERGIZING.

Defective Kilovolt Meter

Check Kilovolt Meter, it may have a defective part.

Содержание 325D

Страница 1: ...MEDICAL SYSTEMS Technical Publication M 2008 7 15 Maintenance Manual For Maintaining and Troubleshooting STANDARD FREQUENCY SERIES GENERATORS Contents Models TXR 325D TXR 325M TXR 425 TXR 525SFQ...

Страница 2: ......

Страница 3: ...sibility of any persons carelessly unwisely or unknowingly exposing themselves or others to X radiation It is important that everyone working with X radiation be properly trained and take adequate ste...

Страница 4: ...ycle of the equipment or system comes to an end becomes dangerous and need to be considered as harmful waste according to the international domestic and local regulations The manufacturer recommends t...

Страница 5: ...ng advisory symbols will be used throughout this manual Their application and meaning are described below DANGER ADVISES OF CONDITIONS OR SITUATIONS THAT IF NOT HEEDED OR AVOIDED WILL CAUSE SERIOUS PE...

Страница 6: ...g documents Ionizing radiation Type B equipment Dangerous voltage Ground This symbol indicates that the waste of electrical and electronic equipment must not be disposed as unsorted municipal waste an...

Страница 7: ...4 4 6 MA Stabilizer 4 5 TROUBLESHOOTING TECHNIQUES 5 5 1 Establishing Troubleshooting Techniques 5 5 1 1 Essentials Required for Producing X ray 5 5 1 2 Three Key Indicators for Effective Troubleshoot...

Страница 8: ...7 8 Troubleshooting Gassy X ray Tube 16 17 7 9 Troubleshooting Low MA 18 7 10 Troubleshooting No MA 18 20 8 COLLIMATOR 21 8 0 Theory of Operation 21 8 1 Mechanical Operation SID Selection 21 8 2 Elec...

Страница 9: ...2008 7 13 M 2008 7 14 and M 2008 7 15 Schematics are located in manual M 2008 7 13 1 1 TOOLS AND TEST EQUIPMENT REQUIRED FOR SERVICING X RAY SYSTEMS The following hand tools and products are required...

Страница 10: ...ements in bold lettering throughout this manual WARNING You are to have proper instrumentation to measure all energies used in and emitted from this equipment At no time are you to expose a person to...

Страница 11: ...ched to Bucky ON change the switch to Bucky OFF and retest If exposure meets condition of 5 above trouble is in the bucky circuit 7 If there is still no exposure initiated under condition 6 the proble...

Страница 12: ...of the Transformer so it can be stepped up to high voltage and Rectified for use by the X ray Tube 4 3 HIGH VOLTAGE TRANSFORMER High Voltage Transformers do not regulate current or voltage Its sole pu...

Страница 13: ...se provides the Keys to establishing effective troubleshooting Techniques The indicators are 1 High MA which can cause light images dark images and kicking circuit breakers 2 Low MA which mainly cause...

Страница 14: ...gh voltage secondary Because of loading kVp will decrease at the X Ray tube The bucking voltage selected from the 95 110 VAC supply bucks H1 to H2 Voltage The final Voltage difference is coupled to th...

Страница 15: ...readings are from the X ray Timer s software 6 3 MA METER The MA meter is a simple Digital resistor diode Display that indicates MA stations selected This meter cannot be adjusted 6 4 TESTING HIGH VO...

Страница 16: ...across major to minor line strap Should read 200 278 VAC Control Energizes but LED s are not lit Open fuse or loose fuse holder A1 or A2 With Line to Auto Strapped properly and with the Circuit Breake...

Страница 17: ...sert Check continuity of tube stator You should find continuity between any pair of the three 3 terminals Gently shake Tube Head or look into port to see if X ray Tube is BROKEN If the tube is broken...

Страница 18: ...posure is now possible Replace the EXPOSURE SWITCH TIMER SECTION 4 During Timer Standby continuous 120 VAC is required on Timer Pins J1 2 and J1 4 Place volt meter between those pins and confirm volta...

Страница 19: ...ed to the Bucky Motor Circuit B3 to B4 The Bucky Motor begins to drive and move the x ray filter grid across the front of the receptor for as long as the switch is depressed At some point the x ray fi...

Страница 20: ...en VISUAL INSPECTION Remove Collimator Select large and then small X ray filaments See if each filament lights up during PREP If both filaments light up properly the problem is in another area Disrega...

Страница 21: ...se EXPECTED VOLTAGE MEASUREMENTS WHEN A SHORT CIRCUIT EXISTS 2 Measure voltage from XC to XS or XC to XL With Prep Switch depressed voltage should be 100 200 VAC If it is remove the collimator and obs...

Страница 22: ...t is or is not operating properly SIMULATED LOAD TEST 1 Remove P1 P2 power cord from the Primary of the High Voltage Transformer 2 Place a Volt Meter between H1 and H2 terminals on the X Ray Control M...

Страница 23: ...ion of High Tension AC Current The Full Wave Bridge rectifies this current to DC DC current measurements dc milliamperes and are taken from and terminals on the Motherboard After the rectifier AC conn...

Страница 24: ...the tube is exposed above 50 kVp at 50 MA Tubes may conduct properly when cold but become unstable when they warm up There is no one set way to measure how gassy a tube has become If a tube is only s...

Страница 25: ...ake an exposure If MA is present you must replace High Voltage Rectifier Board You must establish the reason the Rectifiers shorted Normally a High Voltage Cable shorted or the X ray Tube became gassy...

Страница 26: ...dropped To test the High Voltage Rectifier Board and wiring remove the transformer from the tank and inspect all wiring and connections using a OHM Meter Open Space Charge Primary Transformer Open sp...

Страница 27: ...and the other on the S or L pin depending on which focus is selected Depress PREP and measure voltage If 15 16 VAC is present filament transformer is OK If voltage is not present check input voltage X...

Страница 28: ...move fuse 4 and measure resistance across the 2500 ohm resistor If the resistor is OPEN replace it NOTE If the 2500 ohm resistor heats up or opens you may have a short from MT 1 to MT 2 to a winding o...

Страница 29: ...are positioned manually with these knobs Back illuminated SID scales indicate all conventional film sizes for both 100 cm and 180 cm SID s as well as 40 inches and 72 inches in SIDs All Eureka Linear...

Страница 30: ...F3 on the power supply board and then to collimator receptical connections C and D 8 2 2 LIGHT FIELD LAMP CIRCUIT Refer to Schematic 70 08372 and Assembly Drawing 70 08371 This drawing is provided in...

Страница 31: ...tor Exposure Switch Prep Switch Earth Ground Earth Ground L1 L2 Connection Bucky Switch Timer Board 50 100 MA Adjustments 150 500 MA Adjustments SCR X Ray Tube Rotor Connections 07 08 09 Backup Contac...

Страница 32: ...se for planning or training BMET s MANUAL COST PER COPY Pre Installation Manual M 2008 7 12 20 00 Installation Manual M 2008 7 13 35 00 Operation Manual M 2008 7 14 25 00 Maintenance Manual M 2008 7 1...

Отзывы: