Page 25 of 75 Pages
6.1
Program 1: Glass
For glass instruments and materials, when the manufacturer recommends
autoclaving at temperatures of 134
C (273
F).
Nominal parameters default settings
Sterilization temperature: 134
C (273
F)
Sterilization time: 3 minutes.
Operations Sequence
Insert water into the chamber.
Heating by actuation of electrical heaters until the sterilization
temperature is reached.
Sterilization temperature is maintained constant during the sterilization
time.
Fast exhaust, steam is exhausted out of the chamber at a fast rate until
pressure decreases to ambient pressure.
Note:
The sterility of instruments processed in unwrapped cycles cannot be
maintained if exposed to non-sterile environment.