Message /
Symbol /
Problem
Failure Description
Corrective Action
Heat Time
Error
This message is displayed if the
system cannot reach the required
temperature, in the chamber, within the
preset time.
Verify that the autoclave is not
overloaded.
Heat Time
Error (Keep)
This message is displayed if the
system cannot reach the required
temperature, in the chamber, during
the optional "Keep Heat" stage, within
the preset time.
Verify that the autoclave is not
overloaded.
Low Pressure
This message is displayed if Chamber
Pressure drops below the sterilization
pressure (134°C = 304 kPa, 121°C =
205 kPa) for 2 seconds during the
sterilization stage.
Perform a new cycle.
Low Pressure
(Cooling)
This message is displayed if the
pressure in chamber does not reach the
preset pressure before initiating the
cooling stage.
Check and fix the compressed
air supply.
High Pressure
This message is displayed if Chamber
Pressure raises 4.2 psi-29 kPa above
sterilization pressure (134°C = 304
kPa, 121°C = 205 kPa) for 2 seconds
during the sterilization stage.
Perform a new cycle.
High Pressure
(Ending)
This message is displayed if the system
cannot reach atmospheric pressure ±
10kPa during the ending stage.
Perform a new cycle.
High Pressure
(Exhaust)
This message is displayed if the
system cannot reach preset pressure
within 10 minutes from the beginning
of the exhaust stage.
Perform a new cycle.
Pressure Time
Error
This message is displayed if the
system cannot reach the required
pressure conditions in the chamber,
after preset time, during the air
removal stage.
Verify that the autoclave is not
overloaded.
RTC Error -
Please Set
Current Date
and Time
This message is displayed in order
to set the date and the time.
Set Current Date And Time.
If the problem persists, call the
technician.
Time Error
This message is displayed if the real
time clock is faulty.
Call the technician.
Door is open
(During the
cycle)
This message is displayed when the
door is open: During the cycle.
Close the door to perform a
new cycle.
Canceled By
User
This message is displayed after the
START/STOP key is pressed and
cycle aborted.
Wait until "cycle failed –
canceled by user" or "cycle
end – canceled by user" is
displayed.