Page 64 of 75 Pages
11.6 Unloading
When the cycle ended
successfully
(including pressing the
START/STOP
key, or any failure, after completing the sterilization
stage) message "Cycle Ended" (and the relevant failure message, if
applicable) is displayed on the screen.
Verify that there is no pressure in the chamber, according to the
reading on the display. Only than you may open the door.
Open the autoclave. (see sec. 10.2 Opening the door)
Warning
To avoid sever injuries from hot steam when opening the door:
It is strictly forbidden to lean on the autoclave.
It is strictly forbidden to place your hand or any part of your
body over the door.
Wear heat-resistant gloves or use the tray handle to remove the load
from the autoclave
On completion of the cycle, the load shall be visual inspected to
ascertain that the load is dry, and that the colour of the sterilization
indicators turned to the required colour.
11.7 Stopping the process and cancelling the ERROR message
It is possible to stop the program while the autoclave is operating.
Pressing the
START/STOP
key at any stage of the process stops the
operation. If the cycle was aborted before completing the sterilization
stage, it will leave the load unsterilized.
At the end of the aborted process (before completing the sterilization
stage),
"Cycle Failed"
message, error message and a warning symbol
are displayed on the screen. Refer to "Displayed Error
Messages/Symbols".
Pressing the
START/STOP
key cancels the displayed message and
enables opening the door.
Warning
The load has not completed a sterilization cycle, therefore it is not
sterile. Handle it as contaminated load.
11.8 Start Cycle by Clock mode
This mode enables the operator to define the time of the beginning of
the cycle. The maximum possible delay is 24 hours. For further
information see sec. 6.8.1. “Start Cycle by Clock”.