Page 11 of 75 Pages
3.7
Construction
Chamber and door material
Stainless steel
Outer Cabinet
Stainless Steel
Chamber insulation
Fiberglass with reinforced material
3.8
Loading Capacities
3.8.1
Solid load
Maximum solid load: 30 kg (66.1 lb)
3.8.2
Erlenmeyer Flasks
Size
250 ml
500 ml
1000 ml
Qty
3840
2 x 12
1 x 8
1 x 5
3850
2 x 12
2 x 8
1 x 5
3870
3 x 12
3 x 8
2 x 5
3.8.3
Medium Flasks (Schott)
Size
250 ml.
500 ml
1000 ml
Qty
3840
2 x 19
2 x 12
1 x 8
3850
2 x 19
2 x 12
1 x 8
3870
3 x 19
3 x 12
2 x 8
3.9
Utilities
3.9.1
Electrical utility
Power
Switch box
Recommended
Circuit Breaker
208-220V (3 Ph + Earthing)
No
20A
380-400V(3 Ph + N
Earthing)
No
16A
230V 1ph
yes
32A
Frequency
50/60Hz
3.9.2
Other Utilities
Compressed Air (ELVC only)
1/2" 3-4 Bar (44-58 psi)
Tap water
1/2", 2-3 Bar (29-44 psi)
Mineral free water
1/2", 2-3 Bar (29-44 psi)
Drain
2" Minimum Withstanding temp.
of 80°C (176°F)
Содержание 3840 ELVG-D