Page 55 of 75 Pages
PRINTER OUTPUT
DESCRIPTION
Operator
:
____________
To be filled in manually by operator
Time:
12:14:47
Time sterilization cycle ended
Cycle Ended
00:24:43
101.3
099.7
Cycle finished time
E 00:23:43
107.0
107.4
The time, temperature and pressure during exhaust
E 00:22:08
134.5
311.9
The time, temperature and pressure during exhaust
CLK 2:
12:12:10:00
CLK 1:
12:12:10:00
S 00:22:07
134.5
311.6
The time, temperature and pressure during sterilization
S 00:22:06
134.5
311.6
The time, temperature and pressure during sterilization
S 00:21:06
134.6
311.0
The time, temperature and pressure during sterilization
S 00:20:06
134.5
310.1
The time, temperature and pressure during sterilization
S 00:19:06
134.8
311.1
The time, temperature and pressure during sterilization
S 00:18:06
134.5
315.8
The time, temperature and pressure during sterilization
CLK 2:
12:08:08:00
CLK 1:
12:08:08:00
H 00:18:04
134.4
315.1
The time, temperature and pressure during heating
H 00:16:35
128.9
268.4
The time, temperature and pressure during heating
H 00:13:35
116.3
180.9
The time, temperature and pressure during heating
A 00:11:04
107.0
120.9
The time, temperature and pressure during Air removal
A 00:10:22
115.1
180.4
The time, temperature and pressure during Air removal
A 00:10:00
113.7
170.5
The time, temperature and pressure during Air removal
A 00:07:00
098.7
101.5
The time, temperature and pressure during Air removal
A 00:06:45
097.6
101.4
The time, temperature and pressure during Air removal
A 00:03:45
080.2
099.4
The time, temperature and pressure during Air removal
A 00:00:45
053.7
099.4
The time, temperature and pressure during Air removal
A 00:00:04
046.5
100.0
The time, temperature and pressure during Air removal
TIME
°C
kPa
Ster Time:
4.0min
Sterilization time for selected program
Ster Temp:
134.0°C
Sterilization temperature in chamber for selected program
Cycle:
Unwrapped instru
Cycle name
Cycle Num:
000001
Cycle number
Version:
7.0.00.00
Software version A.B.CC.DD = 7.0.00.00
A: Door Type: Single motor = 7
B: Vacuum Type = 0
C: Total number of Input/Output functionality that are not as default = 00
D: Total number of parameters values that are not as default = 00
Model:
3870ELVC
Model name
Ser. Num:
000000000001
Model Serial number
Time:
11:50:05
Time sterilization cycle started
Date:
9/FEB/2010
Date sterilization cycle started
-----------------------
Time:
08:51:39
Time of turning on
Date:
9/FEB/2010
Date of turning on
POWER ON
The device is turned on
Time:
00:00:00
Time of turning off
Date:
9/FEB/2010
Date of turning off
POWER OFF
The device is turned off
Legend
A
Air removal stage (Steam flash)
E
Exhaust stage
H
Heating stage
S
Sterilization stage
CLK 1
Real Time Clock
C
Cooling stage (for ELVC only)
CLK 2
Software clock