Table of parameters
The parameter default depends on the selected program. The following
table is an example to Custom 1 program
Parameters
Description
Default Unit Range
Resolution
Pulse A Count
Define the number of pulses in the first pulses group
3
#
0-10
1
Pulse A Stay
Time
Define the additional time that the vacuum pump will continue
to operate or slow/top exhaust valve will remain open after
reaching the preset pressure in the first pulses group
10
Second
1-100
1
Pulse A Low
Pressure
Define the minimum vacuum/pressure level in the first pulses
group
130.0
kPa
5-200
1
Pulse A High
Pressure
Define the maximum vacuum/pressure level in the first pulses
group
180.0
kPa
5-200
1
Pulses B Count Define the number of pulses in the second pulses group
0
#
0-10
1
Pulse B Stay
Time
Define the additional time that the vacuum pump will continue
to operate or slow/top exhaust valve will remain open after
reaching the preset pressure in the second pulses group
2
Second
1-100
1
Pulse B Low
Pressure
Define the minimum vacuum/pressure level in the second pulses
group
160.0
kPa
5-200
1
Pulse B High
Pressure
Define the maximum vacuum/pressure level in the second pulses
group
180.0
kPa
5-200
1
Sterilization
Temperature
Define the sterilization temperature
134.0
°C/°F
80-137
0.5°
Sterilization
Time
Define the sterilization time
4.0
Minute 0-9999
0.5
F0Mode
This parameter calculates the time that the load is exposed to
temperature according to F0 table. The printer prints out the
values before sterilization and at the end of this cycle
0
#
0-1
1
Cool Mode
Defines whether to perform the cooling stage or not. There are 4
options: 0 = no cooling. 1 = forced cooling. 2 = N.A for this
model. 3 = N.A for this model.
0
#
0-3
1
Cool End
Temperature
Define the temperature at the end of the cooling stage
80.0
°C/°F
30-120
1
Cool Exhaust
Rate
Defines exhaust rate at the end of cooling stage
5.0
kPa/sec
1-100
1
exhaust Mode
This parameter enables to choose between FAST EXHAUST=1
to SLOW EXHAUST=2
1
#
1-2
1
Dry heat On 1
Defines the time that the heaters operate during the first drying
period
4
Second
0-120
1
Dry Heat Off 1
Defines the time that the heaters do not operate during the first
drying period
12
Second
0-120
1
Dry First Stage
Time
If steam is produced by heating water in the chamber, it is
possible to define two heating periods during the drying stage.
This parameter defines the first heating period. The second
heating period will begin at the end of the first heating period
and will last until the end of the drying stage.
5
Minute
0-120
1
Dry heat On 2
Defines the time that the heaters operate during the second
drying period
2
Second
0-120
1
Dry Heat Off 2
Defines the time that the heaters do not operate during the
second drying period
20
Second
0-120
1
End
Temperature
Define the temperature at the end of the cycle , the cycle will not
end before reaching the preset temperature
120
°C/°F
30-120
1
Multiple Cycles Define the number of times the cycle repeated
1
#
1-50
1
Multiple Cycles
Gap
Define the interval between repeated cycles in case that Multiple
Cycles parameter is defined
2
Minute
1-60
1
Print Rate All
Defines the printing rate during the cycle except sterilization
stage
3
Minute
1-30
1
Print Rate
Defines the printing rate during the sterilization stage
1
Minute
1-30
1