Page 7 of 77
2
SAFETY INSTRUCTIONS
The autoclave has unique characteristics. Please read and understand the operation
instructions before first operation of the autoclave. The following issues may
require instructions guidance provided by the manufacturer: how to operate the
autoclave, the door safety mechanism, the dangers involved in circumventing
safety means, how to ensure that the door is closed, and how to select a correct
sterilization program.
Make sure that you know where the main power switch is, where the water cut-off
valve is and where the compressed air disconnection valve is located.
Autoclave maintenance is crucial for the correct and efficient function of the
device. We enclose a log booklet that includes maintenance recommendations,
with every device.
The weekly spore test is part of the preventive maintenance plan, along with the
annual validation of the sterilization processes that ensures appropriate
temperature dispersion within the chamber.
Never use the autoclave to sterilize corrosive products, such as: acids, bases and
phenols, volatile compounds or solutions such ethanol, methanol or chloroform
nor radioactive substances.
1.
Never start using a new autoclave or a new steam generator, before the
safety, licensing and authorization department has approved it for use.
2.
All autoclave users must receive training in proper usage from an
experienced employee. Every new employee must undergo a training
period under an experienced employee.
3.
A written procedure must be established for autoclave operation,
including: daily safety tests, seal inspection and door hinge inspection,
smooth action of the closing mechanism, chamber cleaning, prevention of
clogging and preservation from corrosion, what is permitted and what is
prohibited for sterilization and choosing a sterilization program.
4.
Liquids may be sterilized only with the “liquids” program. The container
must be covered but not sealed. Sealed bottles may only be sterilized using
a special program. The bottle must be either Pyrex or a Borosilicate glass
bottle.
5.
When sterilizing plastic materials, make sure that the item can withstand
sterilization temperature. Plastic that melts in the chamber is liable to
cause a great deal of damage.
6.
Individual glass bottles may be placed within an appropriate container that
will be placed on a tray. Never place glass bottles on the floor of the
autoclave. Never fill more than 2/3 of the bottle volume.
7.
On closing the autoclave's door, make sure it is properly locked before
activating.
8.
Before withdrawing trays, wear heat resistant gloves.
9.
Before opening the door, verify that there is no pressure in the chamber.
10.
Open the door slowly to allow steam to escape and wait 5 minutes before
you remove the load. When sterilizing liquids, wait 10 minutes.
11.
Once a month, ensure that the safety valves are functioning, and once
annually a certified tester must conduct pressure chamber safety tests.
Содержание 2840 EL-D
Страница 4: ...Page 3 of 77 14 TROUBLESHOOTING 73 15 SPARE PARTS LIST 77 16 ACCESSORIES 77...
Страница 5: ...Page 4 of 77 TABLE OF CONTENT Cont DRAWINGS PAGE NO FRONT VIEW 17 REAR VIEW 18...
Страница 26: ...Page 25 of 77...
Страница 34: ...Page 33 of 77...
Страница 71: ...Page 70 of 77...
Страница 72: ...Page 71 of 77 Cap Gasket Strainer element Strainer Housing WATER STRAINER...
Страница 77: ......