Page 35 of 77
9
INSTALLATION
9.1
Placing
CAUTION:
The sterilizer is not portable or hand-held equipment; it is a fixed device
so it is forbidden to move it.
The sterilizer must be placed on a rigid and leveled surface. The stand
must be able to withstand the load of the device and loaded material.
1.
Counter top
Able to support a minimum of 70 kg (154 lb).
2.
Counter space
Min. 55cmW x 65cmD (22”W x 25”D) * (see unit dimensions).
When changing the autoclave location the door should remain open for
at least 5 minutes in order to set the atmospheric pressure
9.1.1
Placing the Autoclave
Keep the back and the sides of the autoclave approximately 50
mm (2”) away from the wall to allow ventilation and to facilitate
the device disconnection.
If placed in a cabinet, verify that the rear of the cabinet is open to
allow ventilation.
Insufficient space for ventilation may result in an increase of the
autoclave's temperature that may damage the instrument.
It is recommended that enough space be left around the
autoclave to give a technician access for servicing the machine.
9.1.2
Connections to Utility Supplies
Plug the power cord into the power supply output as specified in
sec. 3.5 (Utilities).
9.2
Lifting and carrying
CAUTION:
Before moving the autoclave, Make sure that the electric cord is
disconnected from the power, and there is no pressure in the chamber
and in the generator.
To avoid injuries, lifting and carrying should be done with at least two
persons or by using a fork-lift or any other mechanical aid.
Do not drop the device!
Содержание 2840 EL-D
Страница 4: ...Page 3 of 77 14 TROUBLESHOOTING 73 15 SPARE PARTS LIST 77 16 ACCESSORIES 77...
Страница 5: ...Page 4 of 77 TABLE OF CONTENT Cont DRAWINGS PAGE NO FRONT VIEW 17 REAR VIEW 18...
Страница 26: ...Page 25 of 77...
Страница 34: ...Page 33 of 77...
Страница 71: ...Page 70 of 77...
Страница 72: ...Page 71 of 77 Cap Gasket Strainer element Strainer Housing WATER STRAINER...
Страница 77: ......