Page 10 of 77
3.2
Storage Conditions
The packed or unpacked autoclave shall be retained in indoor
conditions!
3.3
Operating Conditions
This device is for indoor use only!
The sterilizer should be loaded only with autoclavable material!
The environment shall not exceed an ambient temperature of 40
º
C and a
relative humidity of 85% respectively.
The operation altitude shall not be over 2000 meters (6561 feet) (ambient
pressure shall not be lower than 80 kPa (11.6 psi)).
The autoclave shall not be used in a manner not specified in this
manual!
Do not use the autoclave in the presence of dangerous gases.
Operate the autoclave only in the manner specified in the manual. If
the equipment is used in a manner not specified by the manufacturer,
the protection provided by the equipment may be impaired.
Caution!
Waste water should be brought into the public net in accordance with
the local rules or requirements i.e ONLY NON-HAZARDOUS
LIQUIDS SHALL BE DISPOSED IN PUBLIC SEWAGE!