During the process, the stages of the cycle will be displayed on the screen:
* Display can be activated only by an authorized person.
6.1
Operations Sequence
·
Insert water into the chamber and heat by actuation of electrical heaters
until reaching the high preset air pulse.
·
Heating by actuation of electrical heaters until the sterilization
temperature is reached.
·
Sterilization temperature is maintained constant during the sterilization
time.
For Liquids Only:
·
Slow exhaust; steam is slowly exhausted from the chamber, until it
reaches the required end temperature (approx. 95ºC) and the pressure
equals the atmospheric pressure.
The stages names are as follows:
Start
Insert water
Steam flush
Pulse H
Keep Heat (*option)
Heating
Sterilization
Cooling (Only in liquid programs)
Ending