Maintenance
51
It is important to remove a battery pack from service if signs of
diminished run-time or longer charge time are observed. These are
signs of battery pack aging and poor general battery health. A battery
pack in good health will take near 8 to 9 hours to fully charge. A
battery pack that provides 30% less product run-time than expected is
ready to be removed from service. If a given battery pack provides a
run-time of 24 hours or less for the DustTrak
™ monitor or a run-time
of 17 hours or less for the DustTrak
™ monitor with External Pump,
the battery pack should be removed from service and properly
recycled in accordance with local environmental regulations.
Battery Pack Health Status Check
TSI
®
highly recommends consistently monitoring the health of all
battery packs in possession. This status check is intended to help
with that process and should be performed in addition to the use and
handling instructions listed within this manual. This status check is not
intended as a replacement for any other use and handling instructions
as all instructions are important to follow to maintain battery health
and to get the maximum life from battery packs. It is important to
follow the following process to help to identify a battery pack that is no
longer in good health and needs to be recycled.
Battery Pack Health Check Procedure
1. Check the battery pack visually for any signs of bulging or
abnormal appearance.
2. Connect the battery to the TSI
®
Battery Charger (P/N 801809).
3. Plug in power to the charger.
4.
Keep the battery pack on charge until the charger’s “Ready” LED
light is lit green indicating that the battery is now completely
charged.
5. Disconnect power from the charger.
6. Disconnect the battery pack from the charger.
7. Let the battery pack rest (off of charger) for at least one hour.
Содержание DustTrak 8535
Страница 2: ......
Страница 17: ...Parts Identification and Unpacking 7 Item Part Number Description 801906 Router Mounting Kit ...
Страница 18: ...8 Chapter 1 This page intentionally left blank ...
Страница 40: ...30 Chapter 2 Figure 30 Attach Power Cable to Screw Terminals ...
Страница 68: ...58 Appendix A This page intentionally left blank ...
Страница 69: ......