Operation
45
Checklist for Sampling with the Environmental
Enclosure
Before beginning a sample, check to see that all of the following
conditions are satisfied:
✓
All components are properly installed into the enclosure as
described in the
✓
The DustTrak
™ II/DRX monitor has been zeroed at the
temperature at which it will be sampling (if possible).
✓
The Environmental Enclosure has been put in a place clear of
any obstructions that will affect the flow around the enclosure
(putting it on a tripod is optional).
✓
The Enclosure is not resting directly on the ground (no
standing water).
✓
The DustTrak
™ II/DRX monitor has been set to the appropriate
logging mode.
✓
The Enclosure is locked shut to prevent theft or vandalism
to instrument.
✓
Plans have been made to check for maintenance and data
collection every 24 hours after initial set up, if necessary.
After the completion of monitoring, follow the procedures for
transporting the Environmental Enclosure in the
section of
this manual. This prevents the instrument from becoming damaged.
Communication and Data Acquisition
The DustTrak
™ monitor records data in either the manual sampling
mode or the programmable modes. See the DustTrak
™ II or
DustTrak
™ DRX user manual for more information on setting up the
logging modes. To download the data, attach the supplied USB cable
from the USB pass-through connector to a computer USB port, and
use the appropriate commands within the TrakPro
™ software.
NOTE
In order to download a data file, the DustTrak
™
II monitor
CANNOT
be in a sampling mode.
Содержание DustTrak 8535
Страница 2: ......
Страница 17: ...Parts Identification and Unpacking 7 Item Part Number Description 801906 Router Mounting Kit ...
Страница 18: ...8 Chapter 1 This page intentionally left blank ...
Страница 40: ...30 Chapter 2 Figure 30 Attach Power Cable to Screw Terminals ...
Страница 68: ...58 Appendix A This page intentionally left blank ...
Страница 69: ......