Chapter 3
36
Figure 3-16: P-T
RAK
®
Particle Counter Logging Example
To program this logging example, do the following:
1. Make sure the P-T
RAK
®
particle counter is connected to the computer
and turned on.
2. Select
Logging Setup
from the
Instrument Setup
menu. T
RAK
P
RO
™
software retrieves the current settings for Log Modes 2 and 3 from the
P-T
RAK
®
particle counter and displays them (as shown in the previous
dialog).
3.
Enter the following for Log Mode 2 and 3:
Setting
LOG 2
LOG 3
Channels
Particle Conc.
Particle Conc.
Start Date
11/02/2008
11/03/2008
Start Time
08:00
08:00
Log interval
01:00
05:00
Test length
00:08:00 00:08:00
Number of tests
1
4
Time between tests 00:00:00
00:16:00
4. Select
Send
. The logging instrument is programmed for the mode 2 and
mode 3 protocols.
11/02/2008
11/03/2008