Operation
35
The settings for each separate Log mode must
not
require more than
100% of the logger memory.
Note
: If the percent memory required is greater than the available
memory, T
RAK
P
RO
™
software will not allow you to program
the Log Mode 2 or 3. It will only accept a protocol which is
feasible.
3.
When you have finished defining the parameters for LOG 2 and LOG 3
modes, select
Send
.
4. You can now disconnect the P-T
RAK
®
particle counter
and cable from
the computer.
Sample Protocol for Log Mode 2 or 3
The following steps describe how to program a sample protocol for Log
Mode 2 or 3.
Log Mode 2 sample protocol features:
•
The sample protocol for Log Mode 2 is set to take unattended
particle concentration readings for one day.
•
Logging takes place on 11/02/2008.
•
The logging sample begins at 8:00 a.m.
•
Logging continues for eight hours.
Log Mode 3 sample protocol features:
•
The sample protocol for Log Mode 3 is set up to take unattended
particle concentration readings for four days.
•
Logging first takes place on 11/03/2008.
•
The logging sample begins at 8:00 a.m.
•
Logging continues for eight hours.
•
The instrument is off for 16 hours, and then repeats an eight hour
test for the next three days.
The following graphic gives the appearance of the dialog box displayed in
T
RAK
P
RO
™
software, with above-described logging parameters.