TX6000 Series Operation Manual Ver1.4
75
TINY: Use tiny character to display the measurement result, 5ms/meas.
OFF: Measurement result will not be displayed but can be read from the bus.
Query syntax: MEASlay:FRONt?
Return format: <font><NL^END>
<font> can be the following information:
LARGE
TINY
OFF
7.1.2
FREQuency subsystem commands
The FREQuency subsystem commands are mainly used to set the measurement
frequency of the instrument. The :FREQuency query returns the current
measurement frequency.
<value>
Command syntax: FREQuency MIN
MAX
Where,
<value> NR1, NR2 or NR3 data format followed by Hz, kHz, MHz.
MIN Set the measurement frequency as 50Hz (the minimum frequency of
TX6200 is 50Hz and TX6201 is 20Hz)
MAX Set the measurement frequency as 100kHz (the maximum frequency
of TX6000 is 100kHz and the maximum frequency of TX6200/
TX6201 is 200kHz)
For example: WrtCmd(“FREQ 1KHZ”) Set the frequency as 1000Hz.
Query syntax: FREQuency?
Return format: <NR3><NL^END>
7.1.3
VOLTage subsystem commands
The VOLTage subsystem commands are mainly used to set the measurement
voltage. The VOLTage? query returns the current measurement voltage.
Command syntax:
<value>
VOLTage MIN
MAX
Where,
<value> NR1, NR2 or NR3 data format followed by V.
MIN Set the measurement voltage as 5mV.
MAX Set the measurement voltage as 2V.