TX6000 Series Operation Manual Ver1.4
81
The FUNCtion:IMPedance:RANGe:AUTO command is used to set the automatic
range selection status. The FUNCtion:IMPedance:RANGe:AUTO? query returns
the current range status.
Command syntax:
ON (1)
FUNCtion:IMPedance:RANGe:AUTO OFF (0)
Where,
Character 1 (49) is equal to ON.
Character 0 (48) is equal to OFF.
For example: WrtCmd (“FUNC:IMP:RANG:AUTO ON”) Set the automatic range
as ON.
Query syntax: FUNCtion:IMPedance:RANGe:AUTO?
Return format: <NR1><NL^END>
The FUNCtion:Source MONitor:VIAC command is used to set the voltage monitor
ON or OFF. The FUNCTtion:Source MONitor:VIAC? query returns the current
voltage monitor status.
ON (1)
FUNCtion:SMONitor:VIAC OFF (0)
Where,
Character 1 (49) is equal to ON.
Character 0 (48) is equal to OFF.
For example: WrtCmd (“FUNC:SMON:VIAC ON”) Set the voltage monitor as ON.
Query syntax: FUNCtion:SMONitor:VIAC?
Return format: <NR1><NL^END>
The FUNCtion:DEV<n>:MODE command is used to set the deviation
measurement mode. The FUNCtion:DEV<n>:MODE? query returns the current
deviation measurement mode.
Command syntax:
ABSolute
FUNCtion:DEV<n>:MODE PERCent
OFF
Where,
ABSolute Absolute value deviation display
PERCent Percent deviation display
OFF Real value display
Where, <n> is