TX6000 Series Operation Manual Ver1.4
25
3.5.3
Output impedance
TX6000 provides two output impedances for your choice: 100 Ω and 30 Ω. When
testing inductance, it is necessary to input the same output impedance so as to
make data comparison with other instruments.
Note: When an optional bias board is selected, only 100Ω is available.
Operation steps for setting output resistance
Execute the following operations to set output impedance
1)
Move the cursor to the Rsou zone, the following soft keys will be displayed.
100 Ω
30 Ω
2)
Press 100Ω to select the output impedance as 100Ω. Press 30Ω to select the
output impedance as 30Ω.
3.5.4
Average
The AVERAGE function can calculate the average value of two or more test results.
The average times can be set from 1 to 255 with an increase or decrease of 1.
Operation steps for setting test average times.
1)
Move the cursor to the AVG zone, the following soft keys are displayed.
INCR
(
+
)
This key is used to increase the average times.
DECR
(
-
)
This key is used to decrease the average times.
2)
Use above soft keys to set the average times.
3.5.5
Voltage/Current Level Monitor function
The level monitor function can monitor the real voltage across DUT or real current
through DUT. The monitored voltage value is displayed in Vm zone on
<MEASURE DISP> page while the monitored current value is in Im zone.
———————————————————————————————————————
Note: The correction function can influence the level monitor function, so when
the correction data changes the level monitor value will change. When the
correction is switched between OPEN or SHOR or LOAD, the level monitor value
will be influenced as well
———————————————————————————————————————
Operation steps for setting the level monitor function
Execute the following operation steps to set the level monitor function as ON or
OFF.
1)
Move the cursor to Vm/Im zone, the following soft keys will be displayed.