TX6000 Series Operation Manual Ver1.4
26
ON
OFF
2)
Press ON to set the voltage monitor function as ON while press OFF to set the
voltage monitor function as OFF.
3.5.6
DCR polarity
TX6000 can provide two DC resistance test mode: ALT and FIX. At present, only
ALT mode can be used. ALT mode is positive and negative DC voltage
measurement. FIX is to fix positive voltage measurement. ALT mode
is
conductive to the demagnetization when measuring the DC resistance of
inductors, which makes the test more accurate.
Execute the following operation steps to set the DCR polarity.
1)
Move the cursor to DCR POL zone, the following soft keys will be displayed.
FIX
ALT
2)
Press ALT to select alternate mode. Press FIX to select lock of positive level
mode. At present, only ALT mode can be used.
3.5.7
Trigger Delay
TX6000 trigger delay means the delay time from triggering to test-start. Delay
function can set the trigger delay time. When the list sweep test function is used,
all set delay time will be delayed at each sweep test point. The range of the trigger
delay time can be set from 0s to 60s with 1ms as the resolution. The trigger delay
function is great useful when the instrument is applied in an auto test system.
When the instrument is triggered by HANDLER interface, the trigger delay time
can ensure DUT and test terminal has a reliable contact.
Operation steps for setting the delay function
Execute the following steps to set the measurement delay time.
1)
Move the cursor to the DELAY zone.
2)
Use numeric keys to input delay time. After pressing a numeric key, the
following unit keys will be displayed.
msec
sec
3.5.8
Step delay
TX6000 step delay means the delay time from outputting drive signal to test-start.
Delay function can set the step delay time. The range of the step delay time can
be set from 0s to 60s with 1ms as the resolution. When measuring DCR or Rd