TX6000 Series Operation Manual Ver1.4
57
NOTE: When the sweep open/short correction is used, the point-frequency
correction function should be set as OFF. Refer to Correction operation in this
chapter.
5.6
Setup example of comparator
TX6000 provides complete comparator function. It is convenient for
component measurement and judgment, incoming and outgoing quality
inspection in production line. Standard configuration of HANDLER interface
makes it realize automatic sorting measurement system.
The concept and specific operation of the comparator has been described in
the preceding chapter. Here are two setup examples.
5.6.1
Capacitor sorting
Capacitor type: 0805CG271
Basic requirements:
t
he capacitance is divided into two bins: BIN J and BIN K.
When the capacitance passes and the loss fails, it will be sorted additionally.
Measurement parameters: frequency of 100kHz, level of 1Vrms, SLOW, FAIL
alarm and external trigger.
Sorting parameters: J bin -4.6% ~ +4.8%, K bin -9% ~ +10%, loss tg
δ
<
0.15%.
The parameters needed to be set in this example are listed below:
Primary parameter
(
FUN1
)
Cp
Secondary parameter
(
FUN2
)
D
Frequency
(
FRQ
)
100kHz
Level
(
LEV
)
1V
Speed
(
SPEED
)
SLOW
Switch of auxiliary bin
(
AUX
)
ON
Tolerance Mode of primary parameter
(
MODE
)
%TOL (percentage tolerance mode)
Nominal value
(
NOMINAL
)
270pF
Lower limit of BIN1
(
BIN1 LOW
)
-4.6%
Upper limit of BIN1
(
BIN1 HIGH
)
4.8%
Lower limit of BIN2
(
BIN2 LOW
)
-9%
Upper limit of BIN2
(
BIN2 HIGH
)
10%
Lower limit of secondary parameter
(
2nd LOW
)
0.0000
Lower limit of secondary parameter
(
2nd LOW
)
0.0015
Trigger mode
(
TRIG
)
EXT
(
external
)
Alarm mode
(
CMP ALARM
)
POOR ALARM
(
HIGH LONG
)