Initializing pumps
43
C-Series Product Manual
All speeds refer to defined speeds described in the description for the [S]
command (page 66). Speeds can be changed from default initialization
speeds. Slower initialization speeds may be useful when working with vis-
cous fluids or small I.D. (inner diameter) tubing.
Argument <n2> options:
Argument <n3> options:
Distribution valve ports are numbered 1...X starting in a CCW direction from
the syringe port. This is shown in Figures 5-14 (page 55) and 5-15 (page 56).
Initialize pump without valve (W<n>)
The [W] command initializes the plunger drive but doesn't initialize the
valve drive. The [W] command accepts the same <n1> arguments as the [Z]
command.
Initialize valve only (w<n1,n2>)
The [w] command initializes the valve drive only.
Argument <n1> options:
Argument <n2> options:
For <n1> and <n2> to be valid, a distribution valve must be configured. For a
3-way valve, [U11] must be issued to configure appropriately. If the pump is
configured for a non-distribution valve, arguments <n1> and <n2> are
ignored.
<n2> options
Description
0
Set the input initialization port to 1. (default)
1...X
Set the input initialization port to non-default value. (X = number of
ports on the valve.)
<n3> options
Description
0
Set the output initialization port to X. (default)
1...X
Set the output initialization port to non-default value. (X = number of
ports on the valve.)
<n1> options
Description
0
Set the input initialization port to 1. (default)
1...X
Set the input initialization port to non-default value. (X = number of
ports on the valve.)
<n2> options
Description
0
Valves homed in CW direction. (default)
1
Valves homed in CCW direction.