![TriContinent C24000 Скачать руководство пользователя страница 96](http://html.mh-extra.com/html/tricontinent/c24000/c24000_product-manual_1162143096.webp)
Replacing components
96
Maintenance
Replacing components
Replacing syringe
Before starting
Prime with air to remove all fluid from the syringe.
Remove syringe
Refer to Figure 6-17 for this procedure.
1.
Lower the Syringe Plunger Arm/Holder to ensure sufficient room to
remove the Syringe.
–
C3000 pumps: if the pump is powered off, manually lower the
Plunger by pushing firmly down on the Syringe Plunger Arm/Holder.
If the power is on, this can be done by sending the [A3000R] com-
mand.
–
C24000 pumps: with the power on, send the [A24000R] command.
The Plunger cannot be manually lowered on the C24000 pump due
to the high resolution of the drive system.
2.
Twist the Plunger Lock Screw counterclockwise at least two full turns (do
not completely remove the Plunger Lock Screw).
3.
Turn the Syringe counterclockwise and remove from the valve.
4.
Push the Syringe down over the plunger.
5.
Remove the Syringe with Plunger from Syringe Plunger Arm/Holder.
6.
To install the replacement Syringe, follow the steps in "Install new
syringe" below.
Install new syringe
Note:
The C24000 pumps must be initialized before the syringe can be
installed. This is because these pumps cannot be moved manually. For
C24000 pumps, first go to "Initialize pump" (page 30) and return to this
procedure after initialization has been performed.
Refer to Figure 6-17 during this procedure.
1.
Lower the Syringe Plunger Arm/Holder to ensure sufficient room to
install the Syringe and Plunger.
–
C3000 pumps: if the pump is powered off, manually lower the
Plunger by pushing firmly down on the Syringe Plunger Arm/Holder.
If the power is on, this can be done by sending the command
[A3000R].
–
C24000 pumps: with the power on, send the command [A24000R].
The Plunger cannot be manually lowered on the C24000 pump due
to the high resolution of the drive system.
2.
Make sure the Plunger is fully seated in the Syringe, to ensure the length
is short enough to fit into the available space.