Replacing components
98
Maintenance
Replacing valve
CAUTION:
If you would like to replace your current valve with a valve of a
different kind or type, please first contact Customer Service for assistance.
There are some important factors to consider in order to avoid damaging the
equipment.
Before starting
Prime with air to remove all fluid from the syringe.
To remove the valve
1.
Remove syringe as described on page 96.
2.
Disconnect the tubing from the valve.
3.
Remove the two valve screws used to secure the valve to the panel.
4.
Remove the valve from the pump by pulling directly away from the
pump's front face.
To install the new valve
1.
Align internal valve "D” slot with the flat on the motor shaft.
2.
Gently push and rotate the valve into the correct position to align the
pins with the holes in the panel.
3.
Securely tighten (clockwise) the plunger lock screws.
Figure 6-18. Common valve interfaces