Using EEPROM
80
Operating instructions
Set pump configuration EEPROM parameters (u<n_xxx>)
CAUTION:
For factory use only.
The [u] command will load pump configuration and calibration information
into EEPROM. Some of these parameters (AutoRun and CAN baud rate) can
also be changed with the [U] command.
This command does not require an [R] to execute.
Table 5-10 shows the full range of settings, and Table 5-11 shows the
effect of different settings.
Note:
These parameters are read only on power-up. Thus they will only take
effect when power is cycled.
Table 5-10: Pump configuration set with u<n_xxx>
1
Command
Description
C3000/
C3000MP
factory
default
C24000/
C24000MP
factory
default
u1_XXX
Motor hold current, 0-100% of max.
10
10
u2_XXX
Motor run current, 0-100% of max.
75
50
u3_XXX
Default V max and initialization speed in 100 increments/sec (1...60)
14
56
u4_XXX
Plunger stroke in increments * 100
62
248
u5_X
0 = Don't auto initialize valve.
1 = Automatically initialize valve on power-up.
2 = No valve circuitry installed on PCB.
1
1
u6_X
CAN baud rate
0 = CAN bus disabled
1 = 100K
2 = 125K
3 = 250K
4 = 500k
5 = 1M
1
1
u7_XX
Power-up default backlash in increments, [K] command
10
80
u8_X
Number of increments to jam or stall the plunger into the top of the syringe
during plunger initialization.
10
10
u9_XXX
Power-up default gap volume in increments, [k] command
24
384
u10_IOB
3 way valve position
I = Input valve position (0-2)
Position 0 = 0 deg.
Position 1 = 120 deg.
Position 2 = 240 deg.
O = Output valve position (0-2)
B = Bypass valve position (0-2)
210
210