Initializing pumps
41
C-Series Product Manual
Initialization commands
Initialize pump, set valve output to the right (CW polarity)
(Z<n1,n2,n3>)
The [Z] command initializes the plunger drive and sets a non-distribution
type valve output to the right (as viewed from the front of the pump).
If a distribution valve is configured, <n2> and <n3> select the input and out-
put ports. Valve movement occurs in a CW (clockwise) motion. If a non-dis-
tribution valve is configured, <n2> and <n3> are ignored.
The [Z] command resets the velocity settings ([V], [v] and [c]), and acceler-
ation [L] to power-up defaults. The increment mode [N] is preserved after a
[Z] command. For consistency in initialization, it is recommended that ini-
tialization always be performed in the normal (N0) mode, which is the
default.
Arguments:
Argument <n1> options:
All speeds refer to defined speeds described in the description for the [S]
command (page 66). Speeds can be changed from the standard initializa-
tion speeds. Slower initialization speeds may be useful when working with
viscous fluids or small I.D. (inner diameter) tubing.
<n1>
0...40
Plunger stall force and speed.
<n2>
0...X
Set the initialization input port on distribution valves, numbered
clockwise from syringe port. X = number of ports. (Ignored for non
distribution valves.)
<n3>
0...X
Set the initialization output port on distribution valves, numbered
clockwise from syringe port.X = number of ports. (Ignored for non
distribution valves.)
(n1)
0
Default argument
(n2)
1
Default argument
(n3)
X
Default argument, X = number of ports
<n1> options
Description
0
Initializes at full plunger stall force at speed 11 (default).
Recommended for syringes 1.0 mL and larger.
1
Initializes at half plunger stall force at speed 11.
Recommended for 250
µ
L
and 500
µ
L
syringes.
2
Initializes at one third plunger stall force at speed 11.
Recommended for 50
µ
L
and 100
µ
L
syringes.
3
Initializes at full plunger stall force at speed 16.
4
Initializes at full plunger stall force at speed 18.
5...9
Same as <n> = 0
10...40
Initializes at full plunger force and at speed 10...40