Plunger movements
69
C-Series Product Manual
Set auxiliary outputs (J<n>)
The [J] command sets the three auxiliary TTL outputs high or low.
Arguments:
The C-Series pump provides three TTL outputs on connector JP3 (pins 13,
14, and 15), which correspond to outputs 1, 2, and 3 in the table below. The
states of the outputs form a binary equivalent to the J command argument.
They are controlled by the [J] command as follows:
Argument <n> options:
Set auxiliary outputs based on syringe position (j<pppp><n>)
The [j] command sets the three auxiliary TTL outputs to the specified state
<n> when the actual syringe position is less than or equal to the specified
syringe position <ppppn>.
Arguments:
Arguments for C3000/C3000MP:
Arguments for C24000/C24000MP:
<n> 0...7
(see table below)
(0)
default argument
{0}
power-up default
<n> options
Output 3 (pin 15)
1
1.
(0 = low, Gnd; 1 = high, +5VDC)
Output 2 (pin 14)
Output 1 (pin 13)
0
0
0
0
1
0
0
1
2
0
1
0
3
0
1
1
4
1
0
0
5
1
0
1
6
1
1
0
7
1
1
1
<pppp>
See
below
Position value, where auxiliary outputs will be set when the
syringe position is less than or equal to this value. The position
value is measured from top of stroke, which has a value of
zero. Possible argument values are the same as for the [A]
command.
<n>
0...7
Sets TTL output state. Argument works the same as in the [J]
command.
<pppp>
0...3000
Position value in N0 mode (increments)
<pppp>
0...24000
Position value in N1 or N2 mode (micro-increments)
<n>
0...24000
Position value in N0 mode (increments)
<n>
0...192000
Position value in N1 or N2 mode (micro-increments)