![TriContinent C24000 Скачать руководство пользователя страница 66](http://html.mh-extra.com/html/tricontinent/c24000/c24000_product-manual_1162143066.webp)
Plunger movements
66
Operating instructions
Set cutoff velocity in increments/sec (c<n>)
The [c] command sets the cutoff velocity. The cutoff velocity is the velocity
at which the plunger ends its movement. The plunger will slope down [L]
from the top velocity [V]. The [c] command overwrites the [C] command.
Arguments:
The cutoff velocity [c] must be less than or equal to the top velocity [V]. If
an attempt is made to set the cutoff velocity greater than the top velocity,
the cutoff velocity will be automatically set equal to the top velocity.
Set cutoff in increments (C<n>)
During the last phase of a plunger move, the speed ramps down (as defined
by the programmed slope) toward the cutoff velocity [c]. When the cutoff
increments [C] are specified, the plunger stops at <n> increments before
reaching the cutoff velocity. The total number of increments required by the
plunger movement is maintained by adding <n> increments to the second
phase (top velocity) of the plunger move.
Arguments:
For example, [C5] stops the plunger five increments short of the final veloc-
ity. The total number of increments is automatically maintained by adding
five increments to the second phase of plunger movement. In other words,
the deceleration phase of the move is shortened by five increments and the
constant speed phase is lengthened by five increments.
<n>
1...2700
increments/sec, N0 or N1 mode
<n>
1...21600
increments/sec, N2 mode (micro-increments)
{900} Power-up
default
<n>
0...25
increments, as determined by the N mode
{0} power-up
default
Содержание C24000
Страница 8: ...This page intentionally blank ...
Страница 14: ...This page intentionally blank ...
Страница 142: ...This page intentionally blank ...