4.5 Dynamic Calibration Adjustment Procedure using the ECP-100
Reference: FAR 91.217; FAA Advisory Circular 43-6B
FAR 91.423; FAR 43-Appendix E and F
FAA TSO-C88a, EASA ETSO-C88a, SAE AS8003
This procedure will allow adjustment to the calibration curve of the SSD120-(XX)N-RS5 or
SSD120-(XX)NE-RS5 using the ECP-100 as an aide in matching the digitizer output to a primary
flight altimeter or NIST traceable pressure standard. This procedure differs significantly from the
Span Adjust Procedure
described in
§4.4
. The
Dynamic Calibration Procedure
makes
adjustments to the altitude data stored in the digitizer’s EEPROM. The technician may make
adjustments to the digitizer error curve in 1000 foot increments, over the entire operating range
with the single exception of the –1000 foot mark. The technician will make the adjustments at the 0
or whole altitude mark,
NOT
at the parallel data’s nominal transition point. See
Figure 1.
The
digitizer will automatically adjust the ICAO parallel altitude data to transition 50 feet prior to the 0
mark. (i.e. the digitizer’s ICAO parallel altitude code will transition from 900 to 1000 feet while the
serial altitude data is transmitting 950 feet.)
1. Connect the digitizer, ECP-100, NIST standard or flight altimeter as shown in the
Figure 2
and apply power.
2. Slide
the
ECP-100 CAL. PROGRAM
selector to the leftmost
PROGRAM
position.
3. Apply power to the altitude digitizer and slide the
ECP-100
power switch to the on position.
4. Set the altimeter barometric input to 29.92In.Hg. (1013MB). Adjust the static system
pressure and stabilize at the first altitude to be calibrated. The first possible correction for
Trans-Cal digitizers is at 0 feet. All adjustments to the digitizer calibration curve occur at
1000-foot intervals. Use the
ALTITUDE UP
and
ALTITUDE DOWN
buttons to adjust the
ECP-100
to the current pressure altitude prior to pushing the
INITIATE PROGRAM
pushbutton.
5. Press
the
INITIATE PROGRAM
pushbutton once. THE
ECP-100
will enter a digital
correction into the digitizer’s
EEPROM
at the current pressure altitude.
6. Adjust the input pressure to the next 1000-foot increment and adjust the
ECP-100
to the
next 1000-foot increment and repeat step 5. Continue repeating throughout the operating
range of the altitude digitizer.
7. Exercise the aircraft’s static system over the operating range of the altitude digitizer and,
with increasing and decreasing pressure, verify at a minimum of ten test points that the
altitude digitizer and primary flight altimeter correspond within the ±125 foot tolerance.
Lightly tap the altimeter before each reading to eliminate friction. If correspondence is not
achieved at any test point, the altimeter may require calibration.
8. Verify that the digitizer’s output is disabled when the transponder is not in mode “C”, or
when the “Altitude Disable” switch is in the off position.
NOTE:
If an error is entered into the digitizer, adjust the pressure to the correct altitude and re-enter the correction. To
clear
ALL
the corrections to the digitizer error curve press the
ALTITUDE UP PROGRAM
pushbutton once. Then
press and hold the
ALTITUDE DOWN
button for two seconds.
If the digitizer and flight altimeter are within the ±125-foot requirement then no correction is required.
DO NOT adjust the digitizer high and low potentiometers during or after this procedure.
Содержание SSD120-N-RS5
Страница 17: ...Figure 1 Altitude Digitizer Correspondence Page 17 of 57...
Страница 19: ...Figure 2 Dynamic Calibration Set Up Using the ECP 100 Page 19 of 57...
Страница 50: ...Wiring Harness Diagram Part Number 881404 50...
Страница 51: ...Span Adjust Block Diagram 51...
Страница 52: ...Dynamic Calibration Block Diagram 52...
Страница 53: ...Figure 2 Temperature vs Warm up Time TEMPERATURE Degrees Celsius Seconds 20 TIME 53...
Страница 54: ...Outline Drawing 54...