Value Description
Default
0, 1
no averaging
X
2
averaging of 2 values
3
averaging of 3 values
4...15 ...
16
averaging of 16 values
Connection path,
Packed EPATH
with 8 bit class
0x20
0x23
0x24
0x01
0x30
0x6D
Logical Type
= Class
Class #23
Logical Type
= Instance
Instance #1
Logical Type
=Attr.-ID
Attr.-ID #109
To save this attribute permanently the attribute 112
Accept Parameter
or service
code 0x16
save
must be executed.
7.4.3.34 Attribute 110, Number of Magnets
With this attribute the
Number of Magnets
is specified, with which the measuring
system is to be operated. If the input does not agree with the operated number of
magnets, no position is output and in the “Attribute 111, Position Status 154” page, the
error “No magnet detected” is reported.
Data type
USINT
Access
Set / Get
Lower limit
1
Upper limit
3
Default
1
Value Description
Default
1
1 magnet
X
2
2 magnets
3
3 magnets
Connection path,
Packed EPATH
with 8 bit class
0x20
0x23
0x24
0x01
0x30
0x6E
Logical Type
= Class
Class #23
Logical Type
= Instance
Instance #1
Logical Type
=Attr.-ID
Attr.-ID #110
To save this attribute permanently the attribute 112
Accept Parameter
or service
code 0x16
save
must be executed.