8
Beispiel:
Messumformer Typ 2, 1400-0.001mbar
(vgl. Abschnitt 4.2.3)
Istwert: 0.5
mbar
Dekaden: 6
exp0: -3
(f.
1.0e-3mbar)
V
out
=
(10V / 6) x ( log(0.5) - (-3) ) =
4.50 V ;
Stetiger Steuerausgang:
Stellgröße: 0 V entspricht 0%
10 V entspricht 100%
Example:
transducer type 2, 1400-0.001mbar (s.
chapter 4.2.3)
actual pressure:
0.5mbar
decades:
6
exp0:
-3 (f. 1.0e-3mbar)
V
out
=
(10V / 6) x ( log(0.5) - (-3) ) =
4.50 V ;
Steady control output:
regulation variable: 0 V equals 0%
10 V equals 100%
Exemple:
Convertisseur de mesure type 2,
1400-0.001mbar (p. 4.2.3)
pression réelle:
0.5mbar
décades:
6
exp0: -3
(f.
1.0e-3mbar)
V
out
=
(10V / 6) x ( log(0.5) - (-3) ) =
4.50 V ;
Sortie de commande continue:
Grandeur réglante 0 V égale 0 %
10 V égale 100 %