ORIC
®
Piezo Stage Controller PDXC
Chapter 6: Command Line Operation
Page 38
CTN015391-D02
of mm, and Two-Position-Switching mode with rising/falling
for each position, assigned by pos1 and pos2.
25
Get Status of Current
Mode in External
Trigger
ETM?
Returns the current status of external trigger mode,
"ML"(Manual mode), "AR/AF"(Analog-In mode with
rising/falling edge), "FR/FF[value]"(Fixed-Step size mode),
and "PR/PF[pos1],PR/PF[pos2]"(Two-Position Switching
mode).
26
Set External Trigger
Mode
ETM="XXX"
Where "XXX"="ML" means Manual mode (default),
"XXX"="AR/AF" means Analog-In mode with rising/falling
edge set, "XXX"=FR/FF[value] means Fixed-Size mode with
rising/falling edges set and value ranging from -10.0 to
+10.0, "XXX"=PR/PF[pos1],PR/PF[pos2] means Two-
Position-Switching mode with rising/falling for each position,
assigned by pos1 and pos2.
27
Position Limit
PLT=[Min,Max]
Where Min/Max stands for the minimum and maximum
values of position respectively. The controller will not
respond to a value beyond this range.
28
Get Status Range of
Position Limit
PLT?
Returns values of position limit by format of PLT=[Min,Max].
29
Get the Status of
Abnormal Move
Detection.
AMD?
Returns '1' when enabled, '0' when disabled.
30
Set Abnormal Move
Detection
AMD=x
Switches the abnormal move detection on or off, default is
on. Used to detect stage stuck or move by external force,
where x=1 means Enable Detection, while x=0 means
Disable Detection.
31
Get Loop Status
LP?
Returns
‘1’ when open loop, ‘0’ when closed loop.
32
Switch Between
Closed-Loop and
Open-Loop by
Command
LP=x
Where x=1 means open loop (default), while x=0 means
closed loop.
33
Set Step Size for
Open-Loop for
Channel 1
SS=x
Where x ranges from 1 to 65535.
34
Get Step Size for
Open-Loop for
Channel 1
SS?
Return value's range is from 1 to 65535.
35
Set Step Size for
Open-Loop for
Channel 2
SS2=x
Where x ranges from 1 to 65535.
36
Get Step Size for
Open-Loop for
Channel 2
SS2?
Return value's range is from 1 to 65535.
37
Set the Step
Distance in Closed-
Loop Control
SD=x
Where x means the practical distance move each time, in
units of 10 nm. It can only be positive.
38
Set the Step
Distance in Closed-
Loop Control
SD?
Return integers in units of 10 nm, e.g. 100000 stands for 1
mm.
39
Get Error Message
ERR?
Returns the error message (0 means no errors occurred, 1
means command not defined, 2 means command data is
out-of-range, 3 means device failed to execute last
command, 4 means no waveform data loaded, 5 means
stage needs Home first, 6 means device works in the wrong
mode, 7 means stage move abnormal, 8 means excessive