Chapter 4: Panel Operation
Rev B, September 23, 2021
Page 21
Figure 4.20 Save Parameters in SMC Output Mode
4.4. Over Current Protection
The PDXC uses an independent chip to calculate the actual power dissipation of the internal amplifier to ensure
safe operation. When the power dissipation limit is hit or a short circuit happens, the PDXC will disable the output
immediately and display an error message on the screen as shown in Figure 4.21. A power cycle is needed to clear
this error message.
Figure 4.21 Warning Screen
4.5. Firmware Update
The PDXC allows users to update its firmware. The most up-to-date firmware and the firmware update tools are
both available on the software tab of the PDXC page on the Thorlabs website (
To update the controller firmware, connect the controller to a PC with the USB cable included. Keep holding down
the BACK button when powering on the device, then the device will enter the bootloader program designated for
firmware updates.
Figure 4.22 Main Screen of Bootload Program
Press “ENTER” to enter the menu screen
and turn knob to toggle down the screen
until it shows “SAVE SETTINGS?”
SMC mode main screen
Press “ENTER” to save all operation
parameters in SMC mode