Page 53
ETN051885-D02
KIM001 K-Cube Piezo Motor Controller
Fig. 6.6 Position Steps Triggering
Example for a move from 0 to 2000 steps and back.
In forward direction: The first trigger pulse occurs at 1000 steps (StartPosFwd), the
next trigger pulse occurs after another 500 steps (PosIntervalFwd), the stage then
moves to 2000 steps.
In reverse direction: The next trigger occurs when the stage gets to 1200 steps.
Note that position triggering can only be used on one TRIG port at a time.
Trig Polarity
The polarity of the trigger pulse is specified in the
Trig. 1 Polarity
and
Trig 2 Polarity
parameters as follows:
Active is High
- The active state of the trigger port is logic HIGH 5V (trigger input
and output on a rising edge).
Active is Low
- The active state of the trigger port is logic LOW 0V (trigger input
and output on a falling edge).
Note
In the
Trig Out Pos. Steps Fwd
mode, the
StartPos. Fwd
value must be reached in
the forward direction for the subsequent trigger sequence to be performed correctly.
If
StartPos. Fwd
is reached in the reverse direction, the initial trigger will still be
activated, but any subsequent sequence could be corrupted. Similarly,
StartPosRev
must be reached in reverse direction for the sequence to work correctly.
Step Count
Time
Time
Trig Voltage
Pos1 Fwd
Pos2 Fwd
Pos1 Rev
1000
1500
1200
5V
0V
StartPosFwd
StartPosRev
PosIntervalFwd