Page 45
ETN051885-D02
KIM001 K-Cube Piezo Motor Controller
Stop
- halts the movement of the motor.
Settings display
- shows the following user specified settings:
Driver -
the type of control unit associated with the GUI panel.
Drive Voltage
- the maximum piezo drive voltage. Set in the range 85V to 125V.
Drive Rate
- The velocity to move when a move command is initiated. It is specified in
drive pulses/sec, in the range 1 to 2,000.
Fwd Jog Step Size
- the size of step (ie. number of drive pulses) taken when the jog
Forward signal is initiated.
Rev Jog Step Size
- the size of step (ie. number of drive pulses) taken when the jog
Backwards signal is initiated.
Jog Rate
- The velocity to move when a jog command is initiated. It is specified in drive
pulses/sec, in the range 1 to 2,000.
Ident
- when this button is pressed, the text ‘KIM001’ on the top panel display of the
associated hardware unit will flash for a short period.
Active
- lit when the unit is operating normally and a motor is moving.
Error
- lit when a fault condition occurs.
Settings button
- Displays the 'Settings' panel, which allows the operating
parameters to be entered for the motor drive - see Section 6.3.
6.3 Settings Panel
When the 'Settings' button on the GUI panel is clicked, the 'Settings' window is
displayed. This panel allows motor operation parameters such as move/jog velocities
to be modified. Note that all of these parameters have programmable equivalents
accessible through the ActiveX methods and properties on this Control (refer to the
APTServer helpfile
for further details and to Section 1.3.4. for an overview of the APT
ActiveX controls). The various parameters are described as follows.