Rev A Feb 2020
Page 12
Chapter 3 Installation
POWER -
A Standard 3.5 mm front panel jack connector for connecting the unit to a
regulated DC power supply of 15 V, 1A. See Section 1.2. for details on power supply
options.
USB
- USB port for system communications.
ON
- Power ON/Standby switch. When in the ON position, the unit is fully powered
up. When the switch is turned to the Standby position, the unit initiates a controlled
power down sequence, saving all user-adjustable parameters to non-volatile memory
before turning off the power. For the first few seconds, the shutdown can be cancelled
by turning the switch on again, in which case the unit will save the parameters but will
remain powered up. In a powered down (Standby) state, the logic circuits are
powered off and the unit will draw only a small quiescent current. The switch should
always be used to power down the unit.
I/O 1 and I/O 2
- SMA connectors for use with external trigger input and output signals
(5V TTL levels). Units are shipped with I/O functionality disabled. The function is set
to trigger IN or OUT via the GUI settings panel - see Section 6.3.
Warning: Risk of Electrical Shock
The unit must be connected only to a DC supply of 15V, 1A regulated.
Connection to a supply of a different rating may cause damage to the unit and
could result in injury to the operator.
Caution
Do not use a power supply with damaged, cracked or split wires. In such cases,
the supply could still power up the cube, but the integrity of the ground return
may be compromised. This could result in a hazardous live voltage being
generated.
Any obvious signs of damage to the power supply lead should result in
immediate replacement of the PSU.
Note
The USB cable length should be no more than 3 metres unless a powered USB hub
is being used.
Caution
Do not connect the piezo unit to a 'live' external power supply. Doing so (i.e. “hot
plugging”) carries the risk of PERMANENT damage to the unit. Always ensure
the power supply unit is isolated from the mains and the front panel switch turn
OFF before connecting to the piezo unit.
Always power up the piezo unit by connecting its power supply to the mains
then switch ON using the front panel switch. Similarly, to power down the unit,
turn the front panel switch OFF, then disconnect the power supply from the
mains before disconnecting the piezo unit.