ITC4000 Series
Error
Description
Value minimum/maximum reached
Occurs when the range limits of the set points are
reached
Operation is not applicable
Occurs in Edit Name menu if characters can't be de-
leted (with backspace) anymore
FPGA configuration error
A FPGA configuration failure has occurred during the
boot procedure, please switch the unit off and on
again. If this error is shown again contact the Thorlabs
Hotline
Nonvolatile checksum error
A general EEPROM checksum error has occured,
please switch the unit off and on again. If this error will
be appeared again contact the Thorlabs Hotline
General keyboard bootloader error
The keyboard bootloader is not running after the boot
procedure, please switch the unit off and on again. If
this error is shown again contact the Thorlabs Hotline
Hardware error
A device internal hardware error has occurred, please
switch the unit off and on again. If this error continues
contact the Thorlabs Hotline
Calibration memory lost
The calibration data in the EEPROM is invalid, please
switch the unit off and on again. If this error is still
present contact the Thorlabs Hotline
Save/recall memory lost
The saved user setup data in the EEPROM is invalid,
please switch the unit off and on again. If this error is
shown again contact the Thorlabs Hotline
Configuration memory lost
The last saved setup before power down is invalid,
please switch the unit off and on again. If this error is
still present contact the Thorlabs Hotline
Self test failed
The device self test procedure has failed, please
switch the unit off and on again. If this error continues
contact the Thorlabs Hotline