A further way of approximating the relation between temperature and thermistor resistance is
the method according to Steinhart-Hart, described by the formula
with the Steinhart-Hart parameters A, B and C; T - temperature in K. To select the calculation
method and to enter the thermistor parameters, please refer to section
.
3.1.9.2 Current Temperature Sensors AD590 or AD592
If a temperature/current transducer AD590 or AD592 is used as temper-
ature sensor, it has to be connected to pin 10 (-) and pin 11 (+) of the
17W2 mixed DSUB jack "TEC OUTPUT" (R16,
) at the rear of the unit.
3.1.9.3 Voltage Temperature Sensor LM35
If the temperature/voltage transducer LM35 is used as temperature
sensor, it must be connected to pin 10 (out), pin 11 (+) and pin 12 (-)
of the 17W2 mixed DSUB jack "TEC OUTPUT" (R16,
) at the rear of the unit.
3.1.9.4 Voltage Temperature Sensors LM335, LM235 or LM135
If the temperature/voltage transducer LM135, LM235 or LM335 is
used as temperature sensor, it has to be connected to pin 10 (+),
pin 11 (also +) and pin 12 (-) of the 17W2 mixed DSUB jack "TEC
OUTPUT" (R16,
Operating elements at the rear panel
) at the
rear of the unit.
3.1.9.5 Platinum RTD Temperature Sensors PT100 or PT1000
Pin 3 and 4 of the DSUB Connector are the current source, Pin
10 and 11 are the voltage measurement pins for the 4-wire
measurement setup.
One connector of the sensor has to be connected to pin 4 and 11
and the other to pin 3 and 10 of the 17W2 mixed DSUB jack
"TEC OUTPUT" (R16,
Operating elements at the rear panel
)
at the rear of the unit.
For optimum performance, use a separate wire for each pin of
the DSUB connector and connect the corresponding wires only at
the PT100 or PT1000, as close as possible to the sensor.
40
12
12
12
12
Содержание ITC4000 Series
Страница 1: ...Laser Diode Current and Temperature Controller ITC4000 Series Operation Manual 2018 ...
Страница 2: ...Copyright 2018 Thorlabs Version Date Item No 3 9 10 Jul 2018 M0009 510 312 ...
Страница 91: ... 2018 Thorlabs 6 Appendix 89 Set Value Factory Setting Line Frequency Auto Update Capability Disabled ...
Страница 105: ...www thorlabs com ...