Higher values will accelerate the settling to the set temperature. Too high values will increase
the amplitude and number of overshoots or will even make the system instable. The I
parameter is set correctly when the actual temperature reaches the set temperature in a short
time with a minimum of overshoots.
The following table summarizes the effects of increasing the P, D and I share:
Effects of increasing parameters
Parameter
Rise Time
Overshoot
Settling Time
Steady-state Er-
ror
Proportional K
P
Decrease
Increase
Small Change
Decrease
Derivative K
D
Small Decrease
Decrease
Decrease
None
Integral K
I
Decrease
Increase
Increase
Eliminate