© 2018 Thorlabs
6 Appendix
63
6 Appendix
6.1 Technical Data
All technical data are valid at 23 ± 5°C and 45 ± 15% rel. humidity (non condensing).
6.1.1 ITC4001 Laser Control
ITC4001
Front Panel
Remote Control
Current Control (Constant Current Mode)
Laser Diode Current Range
0 to 1 A
Compliance Voltage
11 V
Setting Resolution
0.1 mA
16 µA
Measurement Resolution
0.1 mA
16 µA
Current accuracy
±(0.1 % r 0.5 mA)
Ripple and Noise, 10 Hz to 10 MHz, rms, typ.
w/o Noise Reduction Filter
with Noise Reduction Filter
< 10 µA
< 3 µA
Drift, 24 h (0 -10 Hz, typ., at Constant Ambient Tem-
perature)
< 100 µA
Temperature Coefficient
LD Current Limit
Setting Range
1 mA to 1 A
Setting Resolution
0.1 mA
16 µA
Accuracy
±(0.12% r 0.8 mA)
Power Monitor Input - Photodiode
Photo Current Measurement Ranges
0 to 2 mA / 0 to 20 mA
Photo Current Measurement Resolution
2 mA Range
20 mA Range
1 µA
10 µA
32 nA
320 nA
Photo Current Accuracy,
2 mA Range
20 mA Range
±(0.08% r0.5 µA)
±(0.08% r5 µA)
Photodiode reverse BIAS voltage
0 to 10 V
Photodiode input impedance
~0
W
(Virtual Ground)
Power Monitor Input - Thermopile
Voltage Measurement Ranges
0 to 10 mV / 0 to 100mV / 0 to 1V /
0 to 10V
Voltage Measurement Resolution
1 µV / 10 µV / 100 µV /
1mV
0.16 µV / 1.6 µV / 16 µV /
160 µV
Voltage Measurement Accuracy
±(0.1% r 10 µV / + 100 µV / + 1 mV / + 5 mV)
Voltage Input Impedance
1 M
W
Laser Power Control (Constant Power Mode)
Photo Current Setting Resolution,
Control Range 0 to 2 mA
Control Range 0 to 20 mA
1 µA
10 µA
32 nA
320 nA
Содержание ITC4000 Series
Страница 1: ...Laser Diode Current and Temperature Controller ITC4000 Series Operation Manual 2018 ...
Страница 2: ...Copyright 2018 Thorlabs Version Date Item No 3 9 10 Jul 2018 M0009 510 312 ...
Страница 91: ... 2018 Thorlabs 6 Appendix 89 Set Value Factory Setting Line Frequency Auto Update Capability Disabled ...
Страница 105: ...www thorlabs com ...