Error
Description
LD inhibit is active
The Laser output is temporarily switched off until the
status level on "LD ENABLE IN" jack will be switched
to high level
LD temperature protection is active
The Laser output is disabled because of the "LD
ENABLE IN" error and must be switched on manually
Not permitted with photodiode BIAS
on
Operation is not allowed while the photodiode BIAS is
switched on.
Not permitted with QCW mode on
The constant power mode is not allowed if the QCW
mode is switched on, please set the QCW mode to off
before using constant the power mode
Wrong LD source operating mode
The QCW mode is not allowed in constant power
mode, please switch the operating mode to constant
current before using the QCW mode.
Settings conflict
Only one feedback sensor (photodiode or thermopile)
can be switched to the BNC jack "OPT SENSOR IN".
If this message is shown please set the unused feed-
back sensor to the DSUB connector.
Value is not editable
The selected display item (e.g. measurement reading)
can't be adjusted
Not editable with LD output on
Laser current limit, photo current limit, thermopile
voltage limit, and power limit can't be set if the Laser
output is on.
Not permitted with TEC output on
Operation is not allowed while the TEC output is
switched on. This is valid e.g. for the TEC current limit
and the temperature sensor selection.
Wrong TEC source operating mode
Operation is not allowed in current TEC source oper-
ating mode, please check the selected operating
mode in the TEC source setup
PID auto-tune is currently running
Operation is not permitted during a running PID auto-
tune process
PID auto-tune value error
The values of the PID auto-tune process are out of the
specified range, please try a manuel tuning (refer to
section
)
TEC open circuit detected
The connection to the TEC element is broken or the
resistance of the TEC element is too high, the TEC
output cannot be switched on.
Temperature sensor failure
In case of an incorrect wiring to the temperature
sensor or the selected temperature sensor does not
match
TEC cable connection failure
The TEC setup is not connected to the unit correctly
Value is out of range
A sent numerical value in remote mode is out of
range, check the allowed min/max values for the
parameter to be changed
41
Содержание ITC4000 Series
Страница 1: ...Laser Diode Current and Temperature Controller ITC4000 Series Operation Manual 2018 ...
Страница 2: ...Copyright 2018 Thorlabs Version Date Item No 3 9 10 Jul 2018 M0009 510 312 ...
Страница 91: ... 2018 Thorlabs 6 Appendix 89 Set Value Factory Setting Line Frequency Auto Update Capability Disabled ...
Страница 105: ...www thorlabs com ...