© 2018 Thorlabs
72
ITC4000 Series
6.1.4 ITC4005QCL Laser Control
ITC4005QCL
Front Panel
Remote Control
Laser Current Control (Constant Current Mode)
Laser Diode Current Range
0 to 5 A
Compliance Voltage
20 V
Setting Resolution
1 mA
80 µA
Measurement Resolution
1 mA
80 µA
Current Accuracy
±(0.1 % r 2 mA)
Ripple and Noise, 10 Hz to 10 MHz, rms, typ.
w/o Noise Reduction Filter
with Noise Reduction Filter
< 250 µA
< 50 µA
Drift, 24 h (0 -10 Hz, typ., at Constant Ambient
Temperature)
< 300 µA
Temperature Coefficient
LD Current Limit
Setting Range
5 mA to 5 A
Setting Resolution
1 mA
80 µA
Accuracy
±(0.12% r 3 mA)
Power Monitor Input - Photodiode
Photo Current Measurement Ranges
0 to 2 mA / 0 to 20 mA
Photo Current Measurement Resolution,
2 mA Range
20 mA Range
1 µA
10 µA
32 nA
320 nA
Photo Current Accuracy,
2 mA Range
20 mA Range
±(0.08% r0.5 µA)
±(0.08% r5 µA)
Photodiode Reverse BIAS Voltage
0 to 10 V
Photodiode Input Impedance
~0
W
(Virtual Ground)
Power Monitor Input - Thermopile
Voltage Measurement Ranges
0 to 10 mV / 0 to 100mV / 0 to 1V /
0 to 10V
Voltage Measurement Resolution
1 µV / 10 µV / 100 µV /
1mV
0.16 µV / 1.6 µV / 16 µV /
160 µV
Voltage Measurement Accuracy
±(0.1% r 10 µV / + 100 µV / + 1 mV / + 5 mV)
Voltage Input Impedance
1 M
W
Laser Power Control (Constant Power Mode)
Photo Current Setting Resolution,
Control Range 0 to 2 mA
Control Range 0 to 20 mA
1 µA
10 µA
32 nA
320 nA
Thermopile Voltage Setting Resolution,
Control Range 1 µV to 10 mV
Control Range 10 µV to 100 mV
Control Range 100 µV to 1 V
Control Range 1 mV to 10 V
1 µV
10 µV
100 µV
1 mV
0.16 µV
1.6 µV
16 µV
0.16 mV
Содержание ITC4000 Series
Страница 1: ...Laser Diode Current and Temperature Controller ITC4000 Series Operation Manual 2018 ...
Страница 2: ...Copyright 2018 Thorlabs Version Date Item No 3 9 10 Jul 2018 M0009 510 312 ...
Страница 91: ... 2018 Thorlabs 6 Appendix 89 Set Value Factory Setting Line Frequency Auto Update Capability Disabled ...
Страница 105: ...www thorlabs com ...