© 2015 Thorlabs
62
Series 4000 SCPI Programmers Reference Manual
3.12.4 Temperature Protection
Syntax
SENSe{[1]|3}:TEMPerature:PROTection:WINDow[:AMPLitude]
{MIN|MAX|DEF|<temperature>}
SENSe{[1]|3}:TEMPerature:PROTection:WINDow[:AMPLitude]?
[{MIN|MAX|DEF}]
SENSe{[1]|3}:TEMPerature:PROTection:DELay {MIN|MAX|DEF|<seconds>}
SENSe{[1]|3}:TEMPerature:PROTection:DELay? [{MIN|MAX|DEF}]
SENSe{[1]|3}:TEMPerature:PROTection:TRIPed?
Description
The
SENSe{[1]|3}:TEMPerature:PROTection:WINDow[:AMPLitude]
sets the temperature
window width.
The
SENSe{[1]|3}:TEMPerature:PROTection:DELay
sets the temperature window protection
reset delay.
The
SENSe{[1]|3}:TEMPerature:PROTection:TRIPped?
query command returns an ASCII
1
if the window protection is currently active or an ASCII
0
if it’s inactive.
Example
SENS:TEMP:PROT:WIND 1.5K;DEL 5
Sets the temperature window amplitude to ±1.5 Kelvin and the protection reset delay to
5 seconds for a TED4000 Series instrument.
Note
This command is valid for TED4000 and ITC4000 Series instruments only. For TED4000 Series
instruments the command suffix is 1 (can be omitted), for ITC4000 Series instruments the suffix
to protect the LD output of an ITC4000 Series
instrument.