3 Commands by Subsystem
31
3.4.5 Protection Voltage
Syntax
OUTPut[1]:PROTection:VOLTage[:LEVel] {MIN|MAX|DEF|<volts>}
OUTPut[1]:PROTection:VOLTage[:LEVel]? [{MIN|MAX|DEF}]
Description
The
OUTPut[1]:PROTection:VOLTage[:LEVel]
command sets the output voltage protection
(compliance voltage) threshold. If this threshold is reached the output switches off.
Example
OUTP:PROT:VOLT 8.0
Sets the LD output protection voltage to 8V.
OUTP:PROT:VOLT? MAX
Queries the maximum settable LD output protection voltage.
1.01
Typical response.
Note
This command is valid for LDC4000 and ITC4000 Series instruments only.
3.4.6 Temperature Protection Mode
Syntax
OUTPut[1]:PROTection:INTernal[:MODE] {OFF|PROTection|ENABle}
OUTPut[1]:PROTection:INTernal[:MODE]?
Description
The
OUTPut[1]:PROTection:INTernal[:MODE]
command sets the function of the internal
temperature protection circuit.
Parameter
Description
OFF
Temperature protection is disabled and has no effect on the LD output
PROTection
Acts as protection.
If the temperature is out of bounds the LD output is switched off
ENABle
Acts as enable.
If the temperature is out of bounds the LD output is disabled temporarily.
Example
OUTP:PROT:INT ENAB
Sets the temperature protection mode to output-enable mode.
Note
This command is valid for ITC4000 Series instruments only.
See also
for further details.
62
Содержание 4000 Series
Страница 1: ...Series 4000 LDC40xx TED4015 ITC40xx SCPI Programmer s Reference Manual 2015 ...
Страница 2: ...Version Date 3 3 08 Apr 2015 Copyright 2015 Thorlabs ...
Страница 69: ... 2015 Thorlabs 3 Commands by Subsystem 67 Figure FIGURENO Status Structure ...
Страница 92: ... 2015 Thorlabs 90 Series 4000 SCPI Programmers Reference Manual instrument ...