3 Commands by Subsystem
41
SOUR:CURR:LIM? MAX
Queries the maximum settable LD source limit current of a LDC4000 Series or an
ITC4000 Series instrument.
1.01
Typical response.
Note
This command is valid for LDC4000 and ITC4000 Series instruments only.
3.7.3 LD Current Setpoint
Syntax
SOURce[1]:CURRent[:LEVel][:IMMediate][:AMPLitude] {MIN|MAX|<amps>}
SOURce[1]:CURRent[:LEVel][:IMMediate][:AMPLitude]? [{MIN|MAX}]
Description
The
SOURce[1]:CURRent[:LEVel][:IMMediate][:AMPLitude]
command sets the LD current
setpoint in amperes in current source modes.
Example
SOUR:CURR 1.0
Sets the LD source current to 1.0 A on a LDC4000 Series or an ITC4000 Series
instrument.
SOUR:CURR? MAX
Queries the maximum settable LD source current setpoint of a LDC4000 Series or an
ITC4000 Series instrument.
2.01
Typical response.
Note
This command is valid for LDC4000 and ITC4000 Series instruments only.
3.7.4 LD Optical Power
Syntax
SOURce[1]:POWer[:LEVel][:IMMediate][:AMPLitude] {MIN|MAX|<watts>}
SOURce[1]:POWer[:LEVel][:IMMediate][:AMPLitude]? [{MIN|MAX}]
SOURce[1]:POWer:ALC:SOURce {DIODe|PMETer}
SOURce[1]:POWer:ALC:SOURce?
SOURce[1]:POWer:ALC:BANDwidth {MIN|MAX|DEF|<hertz>}
SOURce[1]:POWer:ALC:BANDwidth? [{MIN|MAX|DEF}]
SOURce[1]:POWer[:LEVel]:DIODe[:CURRent][:IMMediate]
[:AMPLitude] {MIN|MAX|<amps>}
SOURce[1]:POWer[:LEVel]:DIODe[:CURRent][:IMMediate]
[:AMPLitude]? [{MIN|MAX}]
SOURce[1]:POWer[:LEVel]:PMETer[:VOLTage][:IMMediate]
[:AMPLitude] {MIN|MAX|<volts>}
SOURce[1]:POWer[:LEVel]:PMETer[:VOLTage][:IMMediate]
[:AMPLitude]? [{MIN|MAX}]
Содержание 4000 Series
Страница 1: ...Series 4000 LDC40xx TED4015 ITC40xx SCPI Programmer s Reference Manual 2015 ...
Страница 2: ...Version Date 3 3 08 Apr 2015 Copyright 2015 Thorlabs ...
Страница 69: ... 2015 Thorlabs 3 Commands by Subsystem 67 Figure FIGURENO Status Structure ...
Страница 92: ... 2015 Thorlabs 90 Series 4000 SCPI Programmers Reference Manual instrument ...