© 2015 Thorlabs
26
Series 4000 SCPI Programmers Reference Manual
3.3.5 Read Measurement Data
Syntax
READ?
Description
The
READ?
command performs a previously configured measurement, and then sends the
reading to the instrument output buffer. It has the same effect as
INITiate[:IMMediate]
followed by the
FETCh?
command.
Example
The following program segment configures a TED4000 / ITC4000 Series instrument for
temperature measurements and performs one measurement.
CONF:TEMP
Configure the instrument for temperature measurement.
READ?
Perform a measurement and send the reading to the output buffer.
2.01
Typical response.
3.3.6 Simple Measurement
Syntax
For TED4000 Series instruments:
MEASure[:SCALar][:TEMPerature]?
MEASure[:SCALar][:CURRent][1][:DC]?
MEASure[:SCALar]:VOLTage[1][:DC]?
MEASure[:SCALar]:POWer[1]?
MEASure[:SCALar]:TSENsor?
For LDC4000 Series instruments:
MEASure[:SCALar][:CURRent][1][:DC]?
MEASure[:SCALar]:VOLTage[1][:DC]?
MEASure[:SCALar]:CURRent2[:DC]?
MEASure[:SCALar]:POWer2?
MEASure[:SCALar]:VOLTage2[:DC]?
MEASure[:SCALar]:POWer3?
MEASure[:SCALar]:POWer[1]?
For ITC4000 Series instruments:
MEASure[:SCALar]:TEMPerature?
MEASure[:SCALar]:CURRent3[:DC]?
MEASure[:SCALar]:VOLTage3[:DC]?
MEASure[:SCALar]:POWer4?
MEASure[:SCALar]:TSENsor?
MEASure[:SCALar][:CURRent[1]][:DC]?
MEASure[:SCALar]:VOLTage[1][:DC]?
MEASure[:SCALar]:CURRent2[:DC]?
MEASure[:SCALar]:POWer2?
Содержание 4000 Series
Страница 1: ...Series 4000 LDC40xx TED4015 ITC40xx SCPI Programmer s Reference Manual 2015 ...
Страница 2: ...Version Date 3 3 08 Apr 2015 Copyright 2015 Thorlabs ...
Страница 69: ... 2015 Thorlabs 3 Commands by Subsystem 67 Figure FIGURENO Status Structure ...
Страница 92: ... 2015 Thorlabs 90 Series 4000 SCPI Programmers Reference Manual instrument ...