Series 4000 SCPI Programmers Reference Manual
3.9.4 Thermopile/Power Meter Power Range
Syntax
SENSe2:POWer[:DC]:RANGe[:UPPer] {MIN|MAX|<watts>}
SENSe2:POWer[:DC]:RANGe[:UPPer]? [{MIN|MAX}]
Description
The
SENSe2:POWer[:DC]:RANGe[:UPPer]
{MIN|MAX|<watts>}
command sets the
thermopile/power meter power range to the most appropriate range for the given value.
Example
SENS2:POW:RANG 0.012
Sets the power range to or above 12 mW depending on the set responsivity of the used
thermopile/power meter.
SENS2:POW:RANG? MAX
Queries the biggest available power range.
1.00
Typical response.
Note
This command is valid for LDC4000 and ITC4000 Series instruments only.
3.9.5 Thermopile/Power Meter Power Protection
Syntax
SENSe2:POWer[:DC]:PROTection[:LEVel] {MIN|MAX|DEF|<watts>}
SENSe2:POWer[:DC]:PROTection[:LEVel]? [{MIN|MAX|DEF}]
SENSe2:POWer[:DC]:PROTection:TRIPped?
Description
The
SENSe2:POWer[:DC]:PROTection[:LEVel]
command sets the power limit value.
The
SENSe2:POWer[:DC]:PROTection:TRIPped?
query command returns an ASCII
1
if the
power limit value is currently reached or an ASCII
0
if the power is currently below that limit.
Example
SENS:POW:PROT 0.511
Sets the power limit to 511.0 mW.
Note
This command is valid for LDC4000 and ITC4000 Series instruments only.
Содержание 4000 Series
Страница 1: ...Series 4000 LDC40xx TED4015 ITC40xx SCPI Programmer s Reference Manual 2015 ...
Страница 2: ...Version Date 3 3 08 Apr 2015 Copyright 2015 Thorlabs ...
Страница 69: ... 2015 Thorlabs 3 Commands by Subsystem 67 Figure FIGURENO Status Structure ...
Страница 92: ... 2015 Thorlabs 90 Series 4000 SCPI Programmers Reference Manual instrument ...