© 2015 Thorlabs
12
Series 4000 SCPI Programmers Reference Manual
3.1.3 *SAV, *RCL, *SDS - Save/Recall Setup
Syntax
*SAV <value>
*RCL <value>
*SDS <value>
Description
Value: 0..7
Memory location 1 ... 8
Use the
*SAV
command to save the present instrument setup configuration in non-volatile
(state-) memory for later recall. The
*RCL
command is used to restore the instrument to the
saved setup configuration. Eight setup configurations can be saved and recalled.
Use the
*SDS
command to initialize the selected (state-) memory with factory default settings.
Your settings will be overwritten. See also chapter
and
appendix
.
Example
*SAV 1
Saves the present instrument setup configuration to memory location 2.
3.1.4 *TST? - Self Test
Syntax
*TST?
Description
Use this query command to perform the instrument self-test routine. The command places the
coded result in the output queue. A returned value of zero (
0
) indicates that the test passed,
other values indicate that the test failed and the error code is placed into the instrument’s error
queue.
3.1.5 *OPC, *OPC? - Operation Complete
Syntax
*OPC
*OPC?
Description
When
*OPC
is sent, the OPC bit in the
Standard Event Register
will be set after all pending
command operations are complete. When
*OPC?
is sent, an ASCII
1
is placed in the output
queue after all pending command operations are complete.
Typically, either one of these commands is sent after the
INITiate
command. The
INITiate
command is used to take the instrument out of idle in order to perform measurements. While
operating within the trigger model layers, many sent commands will not execute. After all
programmed operations are completed, the instrument returns to the idle state at which time all
pending commands (including
*OPC
and/or
*OPC?
) are executed. When executing the OPC bit
in the status structure is set and/or an ASCII
1
is placed in the output queue.
74
86
Содержание 4000 Series
Страница 1: ...Series 4000 LDC40xx TED4015 ITC40xx SCPI Programmer s Reference Manual 2015 ...
Страница 2: ...Version Date 3 3 08 Apr 2015 Copyright 2015 Thorlabs ...
Страница 69: ... 2015 Thorlabs 3 Commands by Subsystem 67 Figure FIGURENO Status Structure ...
Страница 92: ... 2015 Thorlabs 90 Series 4000 SCPI Programmers Reference Manual instrument ...