3 Commands by Subsystem
61
Description
The
SENSe{[1]|3}:TEMPerature:THERmistor:METHod
command selects one of the two
temperature to/from resistance calculation methods (Steinhart-Hart or Exponential method).
The
SENSe{[1]|3}:TEMPerature:THERmistor[:SHH]:A
command sets the parameter A for
the calculation method according to Steinhart-Hart.
The
SENSe{[1]|3}:TEMPerature:THERmistor[:SHH]:B
command sets the parameter B for
the calculation method according to Steinhart-Hart.
The
SENSe{[1]|3}:TEMPerature:THERmistor[:SHH]:C
command sets the parameter C for
the calculation method according to Steinhart-Hart.
The
SENSe{[1]|3}:TEMPerature:THERmistor:EXPonential:R0
command sets the
parameter R
0
(resistance at nominal temperature T
0
) for the exponential calculation method.
The
SENSe{[1]|3}:TEMPerature:THERmistor:EXPonential:T0
command sets the
parameter T
0
(nominal temperature) for the exponential calculation method.
The
SENSe{[1]|3}:TEMPerature:THERmistor:EXPonential:BETA
command sets the
parameter Beta (also known as B-value) for the exponential calculation method.
Example
SENS:TEMP:THER:METH EXP
Sets the calculation method for a thermistor to exponential for a TED4000 Series
instrument.
SENS3:TEMP:THER:EXP:R0 10k;T0 25;BETA 3988
Sets the three thermistor parameters needed (R
0
, T
0
, Beta) to calculate the temperature
from a given resistance for a ITC4000 Series instrument.
Note
This command is valid for TED4000 and ITC4000 Series instruments only. For TED4000 Series
instruments the command suffix is 1 (can be omitted), for ITC4000 Series instruments the suffix
is 3.
3.12.3 Transducer Offset
Syntax
SENSe{[1]|3}:TEMPerature:OFFSet {MIN|MAX|DEF|<temperature>}
SENSe{[1]|3}:TEMPerature:OFFSet? [{MIN|MAX|DEF}]
Description
The
SENSe{[1]|3}:TEMPerature:OFFSet
command sets a temperature offset for a
temperature sensor (usually used to compensate small errors of temperature sensors).
Example
SENS:TEMP:OFFS -0.2
Sets a temperature offset of -0.2K for a TED4000 Series instrument.
Note
This command is valid for TED4000 and ITC4000 Series instruments only. For TED4000 Series
instruments the command suffix is 1 (can be omitted), for ITC4000 Series instruments the suffix
is 3.
Содержание 4000 Series
Страница 1: ...Series 4000 LDC40xx TED4015 ITC40xx SCPI Programmer s Reference Manual 2015 ...
Страница 2: ...Version Date 3 3 08 Apr 2015 Copyright 2015 Thorlabs ...
Страница 69: ... 2015 Thorlabs 3 Commands by Subsystem 67 Figure FIGURENO Status Structure ...
Страница 92: ... 2015 Thorlabs 90 Series 4000 SCPI Programmers Reference Manual instrument ...